FDA Adverse Event Other Summary report: N

WATERBATH GAS STRUT

MDR report key: 1082236 · Received July 25, 2008

Report

Report Number
1932738-2008-00001
Event Type
Other
Date Received
July 25, 2008
Report Date
July 17, 2008
Manufacturer
MEDTEC, INC. (DBA. CIVCO MEDICAL SOLUTIONS)
Product Code
ILJ
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NC, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

MODIFICATION OF DEVICE - BY AUTHORIZED SVC ORGANIZATION: MFG ORGANIZATION OF ORIGINAL DEVICE SUPPLIED ALTERNATIVE REPLACEMENT PART AS DEFINED ON REPLACEMENT PART SPECIFICATION TO PREVENT MATERIAL DETERIORATION. ALTERNATIVE PART WAS END CAP MISSING BRASS INSERT. SEE SCANNED PAGE.

Description of Event or Problem · 1

WHEN REMOVING A MASK, THE LID ON THE WATERBATH FELL ON A THERAPIST ARM, CAUSING A MINOR CUT AND BRUISING. THE THERAPIST WENT TO HEALTH SVCS, BUT NO TREATMENT WAS NEEDED. THE CUT AND BRUISE HEALED FINE. THE GAS STRUT WHICH HOLDS THE WATERBATH LID OPEN FAILED. THE GAS STRUT WAS A REPLACEMENT GAS STRUT INSTALLED BY THE CLINIC. THE CLINIC DOES NOT BELIEVE THIS TO CAUSE DEATH, LIFE THREATENING INJURY OR A SERIOUS INJURY. A REPLACEMENT GAS STRUT WAS SENT TO THE CLINIC.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 WATERBATH GAS STRUT REPLACEMENT WATERBATH GAS STRUT ILJ MEDTEC, INC. (DBA. CIVCO MEDICAL SOLUTIONS) MT-RH-GS-01 M027000

Patients

Seq Age Sex Outcome Treatment
1