FDA Adverse Event
Malfunction
Summary report: N
CENTURY CCU BED
MDR report key: 1082159
·
Received July 25, 2008
Report
- Report Number
- 1824206-2008-02919
- Event Type
- Malfunction
- Date Received
- July 25, 2008
- Date of Event
- June 27, 2008
- Report Date
- June 27, 2008
- Manufacturer
- HILL-ROM RITTER
- Product Code
- FNL
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- NY, US
- Reporter Occupation
- OTHER
Narratives
Description of Event or Problem · 1
THE HILL-ROM FIELD TECHNICIAN REPORTED THAT THE BRAKES WOULD ENGAGE BUT WOULD NOT HOLD. HE REPLACED THE TORQUE TUBE ASSEMBLY AND THE BRAKES FUNCTIONED AS DESIGNED. HE STATED THAT THE BED WAS OUT OF SERVICE AND NO INJURIES WERE REPORTED CONCERNING THE BRAKE FAILURE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | CENTURY CCU BED | AC POWERED HOSPITAL BED | FNL | HILL-ROM RITTER | 894 | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |