FDA Adverse Event
Malfunction
Summary report: N
CENTRA BED
MDR report key: 1082156
·
Received July 25, 2008
Report
- Report Number
- 1824206-2008-02912
- Event Type
- Malfunction
- Date Received
- July 25, 2008
- Date of Event
- June 25, 2008
- Report Date
- June 25, 2008
- Manufacturer
- HILL-ROM RITTER
- Product Code
- FNL
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- VA, US
- Reporter Occupation
- OTHER
Narratives
Description of Event or Problem · 1
THE HILL-ROM FIELD TECH REPORTED THAT THE BRAKES WOULD ENGAGE BUT DID NOT HOLD. HE FOUND THIS WHILE HE WAS PERFORMING PREVENTIVE MAINTENANCE ON THIS BED AND THERE WERE NO REPORTED INJURIES. HE OPENED UP THE BRAKE BLOCK AND FOUND THAT THE BEARINGS WERE MISSING AND THE TOP BLOCK WAS BROKEN. HE REPLACED THE TOP BLOCK AND INSTALLED BEARINGS AND THE BRAKES FUNCTIONED PROPERLY.