FDA Adverse Event
Malfunction
Summary report: N
VERSACARE
MDR report key: 1082127
·
Received July 25, 2008
Report
- Report Number
- 1824206-2008-02890
- Event Type
- Malfunction
- Date Received
- July 25, 2008
- Date of Event
- January 20, 2006
- Report Date
- January 20, 2006
- Manufacturer
- HILL-ROM RITTER
- Product Code
- FNL
- Report Source
- Manufacturer report
- Reporter Location
- ME, US
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
THE PT WHICH WAS PLACED IN THIS BED WEIGHT IS SMALLER THAN THE INTENDED USE OF THE PROD. THE TECH COULD FIND NO PROBLEMS WITH THE BED, AND FOUND THAT THE BED OPERATED WITHIN SPECS.
Description of Event or Problem · 1
A NURSE ALLEGED, SHE FOUND A PT'S WRIST ENTRAPPED BETWEEN THE UNDER SIDE OF THE RIGHT HEAD SIDERAIL AND THE MATTRESS. THE NURSE REPORTED THAT THE BED WAS IN THE LOW POSITION AND THE SLEEP SURFACE WAS FLAT WHEN THE ENTRAPMENT OCCURRED. THE NURSE REPORTED THAT THE PT WAS NOT INJURED DUE TO THE ENTRAPMENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | VERSACARE | AC POWERED HOSPITAL BED | FNL | HILL-ROM RITTER | P3200 | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |