FDA Adverse Event Malfunction Summary report: N

INRATIO

MDR report key: 1082071 · Received July 23, 2008

Report

Report Number
2954730-2008-00474
Event Type
Malfunction
Date Received
July 23, 2008
Date of Event
July 1, 2008
Report Date
July 22, 2008
Manufacturer
HEMOSENSE, INC.
Product Code
JPA
Removal / Correction Number
NI
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
TX, US
Reporter Occupation
UNKNOWN

Narratives

Additional Manufacturer Narrative · 1

DISCREPANT RESULTS (ACCURACY) COMPARISON OF INRATIO TEST WITH LAB RESULTS PROVIDED BY END-USER AT TIME COMPLAINT WAS FILED: DATE: 2008; INRATIO: 4.3; LAB: 2.7; MEAN: 3.5; CONFIDENCE LIMITS: 2.0-5.0; DATE: 2008; INRATIO: 4.8; LAB: 3.8; MEAN: 4.3; CONFIDENCE LIMITS: 2.5-6.3. PER INTERNAL PROCEDURE, THE MEAN OF THE INRATIO METER AND COMPARATIVE SYSTEM INR WERE CALCULATED. BOTH INRATIO AND LAB VALUES ARE WITHIN THE CONFIDENCE LIMITS FOR INR TESTING. THE RESULTS ARE NOT CONSIDERED DISCREPANT WITHIN THE CONTEXT OF THE DOCUMENTED VARIABILITY FOR INR TESTING. THEREFORE FURTHER TESTING IS NOT REQUIRED AT THIS TIME. PRODUCTS WILL BE TESTED WHEN RETURNED.

Description of Event or Problem · 1

CALLER ALLEGED DISCREPANT RESULTS COMPARED WITH THE LAB. RESULTS AS FOLLOWS: DATE: 2008; INRATIO: 4.3; LAB: 2.7. DATE: 2008; INRATIO: 4.8; LAB: 3.8.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 INRATIO PROTHROMBIN TIME TEST JPA HEMOSENSE, INC. 0100007 070761

Patients

Seq Age Sex Outcome Treatment
1 NI