INRATIO
Report
- Report Number
- 2954730-2008-00474
- Event Type
- Malfunction
- Date Received
- July 23, 2008
- Date of Event
- July 1, 2008
- Report Date
- July 22, 2008
- Manufacturer
- HEMOSENSE, INC.
- Product Code
- JPA
- Removal / Correction Number
- NI
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- TX, US
- Reporter Occupation
- UNKNOWN
Narratives
DISCREPANT RESULTS (ACCURACY) COMPARISON OF INRATIO TEST WITH LAB RESULTS PROVIDED BY END-USER AT TIME COMPLAINT WAS FILED: DATE: 2008; INRATIO: 4.3; LAB: 2.7; MEAN: 3.5; CONFIDENCE LIMITS: 2.0-5.0; DATE: 2008; INRATIO: 4.8; LAB: 3.8; MEAN: 4.3; CONFIDENCE LIMITS: 2.5-6.3. PER INTERNAL PROCEDURE, THE MEAN OF THE INRATIO METER AND COMPARATIVE SYSTEM INR WERE CALCULATED. BOTH INRATIO AND LAB VALUES ARE WITHIN THE CONFIDENCE LIMITS FOR INR TESTING. THE RESULTS ARE NOT CONSIDERED DISCREPANT WITHIN THE CONTEXT OF THE DOCUMENTED VARIABILITY FOR INR TESTING. THEREFORE FURTHER TESTING IS NOT REQUIRED AT THIS TIME. PRODUCTS WILL BE TESTED WHEN RETURNED.
CALLER ALLEGED DISCREPANT RESULTS COMPARED WITH THE LAB. RESULTS AS FOLLOWS: DATE: 2008; INRATIO: 4.3; LAB: 2.7. DATE: 2008; INRATIO: 4.8; LAB: 3.8.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | INRATIO | PROTHROMBIN TIME TEST | JPA | HEMOSENSE, INC. | 0100007 | 070761 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | NI |