FDA Adverse Event Malfunction Summary report: N

INRATIO

MDR report key: 1082042 · Received July 23, 2008

Report

Report Number
2954730-2008-00476
Event Type
Malfunction
Date Received
July 23, 2008
Date of Event
June 27, 2008
Report Date
July 23, 2008
Manufacturer
HEMOSENSE, INC.
Product Code
JPA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Occupation
UNKNOWN

Narratives

Additional Manufacturer Narrative · 1

DISCREPANT RESULTS (ACCURACY) COMPARISON OF INRATIO TEST WITH LAB RESULTS PROVIDED BY END-USER AT TIME COMPLAINT WAS FILED: DATE: 2008, INRATIO: 2.8, LAB: 2.6, MEAN: 2.7, CONFIDENCE LIMITS: 1.7-3.8. PER INTERNAL PROCEDURE, TR 0150, THE MEAN OF THE INRATIO METER AND COMPARATIVE SYSTEM INR WERE CALCULATED. BOTH INRATIO AND LAB VALUES ARE WITHIN THE CONFIDENCE LIMITS FOR INR TESTING. THE RESULTS ARE NOT CONSIDERED DISCREPANT WITHIN THE CONTEXT OF THE DOCUMENTED VARIABILITY FOR INR TESTING. THEREFORE, FURTHER TESTING IS NOT REQUIRED AT THIS TIME. PER TEXT "06/26/08 - REPORTER STATED THAT THE RESULTS ARE TOO HIGH, FOR THE PT SECONDS. A DAY PRIOR - MONITOR PT 11.0 VS LAB PT 17.2". CALLER PROVIDED PT VALUES ONLY. INSUFFICIENT INFO AVAILABLE. NO FURTHER TESTING CAN BE PERFORMED.

Description of Event or Problem · 1

CALLER ALLEGED DISCREPANT RESULTS COMPARED WITH THE LAB. RESULTS AS FOLLOWS: DATE: 2008, INRATIO: 2.8, LAB: 2.6.