INRATIO
Report
- Report Number
- 2954730-2008-00476
- Event Type
- Malfunction
- Date Received
- July 23, 2008
- Date of Event
- June 27, 2008
- Report Date
- July 23, 2008
- Manufacturer
- HEMOSENSE, INC.
- Product Code
- JPA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Occupation
- UNKNOWN
Narratives
DISCREPANT RESULTS (ACCURACY) COMPARISON OF INRATIO TEST WITH LAB RESULTS PROVIDED BY END-USER AT TIME COMPLAINT WAS FILED: DATE: 2008, INRATIO: 2.8, LAB: 2.6, MEAN: 2.7, CONFIDENCE LIMITS: 1.7-3.8. PER INTERNAL PROCEDURE, TR 0150, THE MEAN OF THE INRATIO METER AND COMPARATIVE SYSTEM INR WERE CALCULATED. BOTH INRATIO AND LAB VALUES ARE WITHIN THE CONFIDENCE LIMITS FOR INR TESTING. THE RESULTS ARE NOT CONSIDERED DISCREPANT WITHIN THE CONTEXT OF THE DOCUMENTED VARIABILITY FOR INR TESTING. THEREFORE, FURTHER TESTING IS NOT REQUIRED AT THIS TIME. PER TEXT "06/26/08 - REPORTER STATED THAT THE RESULTS ARE TOO HIGH, FOR THE PT SECONDS. A DAY PRIOR - MONITOR PT 11.0 VS LAB PT 17.2". CALLER PROVIDED PT VALUES ONLY. INSUFFICIENT INFO AVAILABLE. NO FURTHER TESTING CAN BE PERFORMED.
CALLER ALLEGED DISCREPANT RESULTS COMPARED WITH THE LAB. RESULTS AS FOLLOWS: DATE: 2008, INRATIO: 2.8, LAB: 2.6.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | INRATIO | PROTHROMBIN TIME TEST | JPA | HEMOSENSE, INC. | 0100004 | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | NI |