FDA Adverse Event Malfunction Summary report: N

CYPHER SIROLIMUS-ELUTING CORONARY STENT

MDR report key: 1081981 · Received July 22, 2008

Report

Report Number
9616099-2008-01816
Event Type
Malfunction
Date Received
July 22, 2008
Report Date
July 11, 2008
Manufacturer
CORDIS DE MEXICO
Product Code
NIQ
PMA / PMN Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
JA
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE STENT SYSTEM IS AVAILABLE FOR EVALUATION AND TESTING; HOWEVER, THE DEVICE HAS NOT BEEN RETURNED AS OF TO DATE (WHICH IS CODED AS "OTHER" ). ANY ADDITIONAL INFORMATION WILL BE SUBMITTED WITHIN 30 DAYS UPON RECEIPT. THIS DEVICE IS NOT DISTRIBUTED IN THE UNITED STATES; HOWEVER, IT IS SIMILAR TO THE CYPHER SIROLIMUS-ELUTING STENTS DISTRIBUTED IN THE US.

Description of Event or Problem · 1

DURING A CORONARY STENTING PROCEDURE, THE CYPHER STENT DELIVERY CATHETER BALLOON RUPTURED. THE TARGET SITE WAS THE MID RIGHT CORONARY ARTERY LESION, WHICH WAS CHARACTERIZED AS DE NOVO, HEAVILY CALCIFIED AND MODERATELY TORTUOUS WITH 100% STENOSIS (CHRONIC TOTAL OCCLUSION). A GUIDEWIRE WAS ABLE TO CROSS THE LESION; HOWEVER A PREDILATATION BALLOON CATHETER WAS UNABLE TO CROSS THE LESION DUE TO THE HEAVY CALCIFICATION. THEREFORE, ROTABLATOR AND INVASCULAR ULTRASOUND (IVUS) WERE CONDUCTED AND PREDILATION WAS COMPLETED WITH A POWERED LACROSSE 2.75X15MM HIGH-PRESSURE BALLOON. THE CYPHER 3.00X33MM STENT WAS DELIVERED TO THE LESION WITHOUT DIFFICULTY; HOWEVER, THE BALLOON COULD NOT BE INFLATED. A BALLOON RUPTURE WAS SUSPECTED, AND THEREFORE, ANOTHER CYPHER STENT WAS USED AND IMPLANTED AT THE TARGET SITE. THE PROCEDURE WAS SUCCESSFULLY COMPLETED WITHOUT INCIDENT TO THE PATIENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 CYPHER SIROLIMUS-ELUTING CORONARY STENT DRUG-ELUTING STENT (NIQ) NIQ CORDIS DE MEXICO NA 13396912

Patients

Seq Age Sex Outcome Treatment
1 UNK IKAZUCHI 1.5/10MM| AND POWERED LACROSSE 2.75/15MM BALLOON CATHETERS| ATHLETE GT SOFT| EVEREST 30 INDEFLATOR| CONQUESTPRO| AND TORNUS GUIDEWIRES