FDA Adverse Event Malfunction Summary report: N

PROGRAMMING WAND

MDR report key: 1081943 · Received July 23, 2008

Report

Report Number
1644487-2008-01691
Event Type
Malfunction
Date Received
July 23, 2008
Date of Event
June 1, 2008
Report Date
June 26, 2008
Manufacturer
CYBERONICS, INC.
Product Code
LYJ
PMA / PMN Number
P970003
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NY, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

DEVICE FAILURE IS SUSPECTED, BUT DID NOT CAUSE OR CONTRIBUTE TO A DEATH OR SERIOUS INJURY.

Description of Event or Problem · 1

REPORTER INDICATED THAT COMMUNICATION COULD NOT BE ESTABLISHED WITH A TEST GENERATOR WHEN USING A SITE'S PROGRAMMING WAND. WHEN ANOTHER HANDHELD COMPUTER WAS ATTACHED TO THE WAND, COMMUNICATION STILL COULD NOT BE ESTABLISHED INDICATING THAT THE WAND WAS A LIKELY SOURCE OF THE PROBLEM. THE WAND BATTERIES WERE CHANGED, CONNECTIONS WERE CHECKED, AND THERE WAS NOT BELIEVED TO BE ANY EMI. THE WAND HAS BEEN RETURNED TO THE MANUFACTURER FOR PRODUCT ANALYSIS, AND A NEW WAND HAS BEEN SENT TO THE SITE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 PROGRAMMING WAND LYJ CYBERONICS, INC. 201

Patients

Seq Age Sex Outcome Treatment
1