FDA Adverse Event Malfunction Summary report: N

BIPOL LEAD MODEL 300

MDR report key: 1081928 · Received July 23, 2008

Report

Report Number
1644487-2008-01664
Event Type
Malfunction
Date Received
July 23, 2008
Date of Event
January 1, 2008
Report Date
June 23, 2008
Manufacturer
CYBERONICS, INC.
Product Code
LYJ
PMA / PMN Number
P970003
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
OH, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

DEVICE FAILURE IS SUSPECTED, BUT DID NOT CAUSE OR CONTRIBUTE TO A DEATH OR SERIOUS INJURY.

Description of Event or Problem · 1

PATIENT UNDERWENT GENERATOR END OF SERVICE GENERATOR REPLACEMENT SURGERY. UPON CONNECTING THE NEW GENERATOR TO THE EXISTING LEAD, SYSTEMS DIAGNOSTICS RESULTED IN HIGH IMPEDANCE. TROUBLESHOOTING EFFORTS TO ENSURE PROPER CONNECTION BETWEEN THE GENERATOR AND LEAD DID NOT RESOLVE THE HIGH IMPEDANCE EVENT AND THE LEAD WAS SUBSEQUENTLY REPLACED. PRODUCT ANALYSIS WILL BE PERFORMED ONCE PRODUCT IS RETURNED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 BIPOL LEAD MODEL 300 LYJ CYBERONICS, INC. 300-20 19227C

Patients

Seq Age Sex Outcome Treatment
1