FDA Adverse Event Malfunction Summary report: N

CONTOUR TEST STRIPS (50)

MDR report key: 1081849 · Received July 18, 2008

Report

Report Number
1826988-2008-00782
Event Type
Malfunction
Date Received
July 18, 2008
Date of Event
June 26, 2008
Report Date
June 26, 2008
Manufacturer
BAYER HEALTHCARE, LLC
Product Code
NBW
Removal / Correction Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
FL, US
Reporter Occupation
OTHER

Narratives

Description of Event or Problem · 1

THE CUSTOMER ALLEGED THAT HE RECEIVED A HIGH BLOOD GLUCOSE READING USING HIS CONTOUR METER. WHILE TROUBLESHOOTING, HE PERFORMED CONTROL TESTS AND RECEIVED A RESULT OF 230 MG/DL. THE NORMAL CONTROL RANGE WAS 102-141 MG/DL. NO ADVERSE EVENTS WERE ALLEGED. THE TEST STRIPS WERE RETURNED FOR EVALUATION. REPLACEMENT TEST STRIPS WERE SENT TO THE CUSTOMER.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 CONTOUR TEST STRIPS (50) BLOOD GLUCOSE TEST STRIPS NBW BAYER HEALTHCARE, LLC 7080D 7LC3C15

Patients

Seq Age Sex Outcome Treatment
1 UNK