GORE TEX VASCULAR GRAFT (TAPERED)
Report
- Report Number
- 2017233-2008-00413
- Event Type
- Injury
- Date Received
- July 25, 2008
- Date of Event
- July 1, 2008
- Report Date
- July 25, 2008
- Manufacturer
- W.L. GORE & ASSOCIATES
- Product Code
- DSY
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- NY, US
- Reporter Occupation
- PHYSICIAN
Narratives
NICHOLAS J. GARGIULO NJ III, VEITH FJ, SCHER LA, MD, LIPSITZ EC, SUGGS WD, BENROS RM. EXPERIENCE WITH COVERED STENTS FOR THE MGMT OF HEMODIALYSIS POLYTETRAFLUOROETHYLENE GRAFT SEROMAS. J VASC SURG 2008;48:216-7.
AS REPORTED VIA THE MEDICAL LITERATURE, A FEMALE UNDERWENT RIGHT UPPER EXTREMITY GRAFTING FOR HEMODIALYSIS FROM THE BRACHIAL ARTERY TO THE AXILLARY VEIN USING A STANDARD WALL 4 TO 7 MM TAPERED POLYTETRAFLUOROETHYLENE (PTFE) (W.L. GORE & ASSOCIATES, INC, FLAGSTAFF, ARIZ) GRAFT. SHE DEVELOPED A GRAFT SEROMA AND WAS MANAGED SUCCESSFULLY WITH OPEN DRAINAGE AND PLACEMENT OF A NO 7 JACKSON PRATT DRAIN. TRANSUDATION OF FLUID WAS SEEN EMANATING FROM THE ARTERIAL END OF THE PTFE GRAFT. SIX WEEKS FOLLOWING REMOVAL OF THE DRAIN, THE SEROMA RECURRED. AN 8 MM X 50 MM WALLGRAFT (BOSTON SCIENTIFIC, NATICK, MASS) WAS DEPLOYED IN THE ARTERIAL PORTION OF THE GRAFT RESULTING IN COMPLETE RESOLUTION OF THE SEROMA.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | GORE TEX VASCULAR GRAFT (TAPERED) | NONE | DSY | W.L. GORE & ASSOCIATES | WLG211 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 62 YR | Required Intervention |