FDA Adverse Event Injury Summary report: N

NXSTAGE SYSTEM ONE

MDR report key: 1081687 · Received July 24, 2008

Report

Report Number
3003464075-2008-00334
Event Type
Injury
Date Received
July 24, 2008
Date of Event
June 27, 2008
Report Date
June 27, 2008
Manufacturer
NXSTAGE MEDICAL, INC.
Product Code
KDI
PMA / PMN Number
K050525
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Occupation
NOT APPLICABLE

Narratives

Additional Manufacturer Narrative · 1

THE REPORTED BLOOD LOSS IS ATTRIBUTED TO CLOTTING OF THE EXTRACORPOREAL BLOOD CIRCUIT. THERE IS NO EVIDENCE OF A DEVICE MALFUNCTION. THE CYCLER ALARMED APPROXIMATELY TO THE CLOTTING. A DIRECT CORRELATION BETWEEN NXSTAGE SYSTEM ONE AND THE REPORTED BLOOD LOSS CANNOT BE ESTABLISHED. FACILITY STAFF HAS BEEN NOTIFIED. NXSTAGE MEDICAL CONSIDERS THIS REPORT CLOSED. NO ADDITIONAL INFO WILL BE PROVIDED.

Description of Event or Problem · 1

THIS REPORT IS BEING FILED AS AN ADVERSE EVENT SOLELY TO COMPLY WITH THE FDA'S BLOOD LOSS POLICY. A VENOUS PRESSURE HIGH ALARM OCCURRED TOWARD THE END OF DURING A ROUTINE HEMODIALYSIS TREATMENT, WHICH COULD NOT BE RESOLVED BY THE OPERATOR. TREATMENT WAS DISCONTINUED WITHOUT PERFORMING RINSEBACK DUE TO CLOTTING OF THE EXTRACORPOREAL CIRCUIT, RESULTING IN AN ESTIMATED BLOOD LOSS OF 190C. NO MEDICAL INTERVENTION WAS REQUIRED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 NXSTAGE SYSTEM ONE HIGH PERMEABILITY HEMODIALYSIS SYSTEM KDI NXSTAGE MEDICAL, INC. CAR-170 8037712

Patients

Seq Age Sex Outcome Treatment
1 44 YR Other