FDA Adverse Event
Injury
Summary report: N
NXSTAGE SYSTEM ONE
MDR report key: 1081684
·
Received July 24, 2008
Report
- Report Number
- 3003464075-2008-00337
- Event Type
- Injury
- Date Received
- July 24, 2008
- Date of Event
- June 30, 2008
- Report Date
- June 30, 2008
- Manufacturer
- NXSTAGE MEDICAL, INC.
- Product Code
- KDI
- PMA / PMN Number
- K050525
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Occupation
- NOT APPLICABLE
Narratives
Additional Manufacturer Narrative · 1
THERE IS NO EVIDENCE OF A DEVICE MALFUNCTION. FACILITY STAFF ATTRIBUTED THE ALARMS TO ACCESS FLOW PROBLEMS. THE CYCLER ALARMED APPROPRIATELY. A DIRECTION CORRELATION BETWEEN NXSTAGE SYSTEM ONE AND THE REPORTED BLOOD LOSS CANNOT BE ESTABLISHED. NXSTAGE MEDICAL CONSIDERS THIS REPORT CLOSED. NO ADDITIONAL INFO WILL BE PROVIDED.
Description of Event or Problem · 1
THIS REPORT IS BEING FILED AS AN ADVERSE EVENT SOLELY TO COMPLY WITH THE FDA'S BLOOD LOSS POLICY. VENOUS AIR ALARMS OCCURRED AFTER SWITCHING PORT CONNECTIONS DURING A ROUTINE HEMODIALYSIS TREATMENT. THE OPERATOR WAS UNABLE TO RESOLVE THE ALARMS AND ENDED TREATMENT. RINSEBACK WAS NOT PERFORMED RESULTING IN AN ESTIMATED BLOOD LOSS OF 190CC. NO MEDICAL INTERVENTION WAS REQUIRED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | NXSTAGE SYSTEM ONE | HIGH PERMEABILITY HEMODIALYSIS SYSTEM | KDI | NXSTAGE MEDICAL, INC. | CAR-170 | 8037702 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 48 YR | Other |