PUMP MMT-522NAS PRDGM INS V2.2 SK EN
MDR report key: 1081653
·
Received July 23, 2008
Report
- Report Number
- 3004209178-2008-00545
- Event Type
- Injury
- Date Received
- July 23, 2008
- Date of Event
- July 14, 2008
- Report Date
- July 15, 2008
- Manufacturer
- MEDTRONIC PUERTO RICO OPERATIONS MED-REL
- Product Code
- LZG
- PMA / PMN Number
- P980022
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- OH, US
- Reporter Occupation
- NOT APPLICABLE
Narratives
Additional Manufacturer Narrative · 1
CURRENTLY, IT IS UNKNOWN WHETHER OR NOT THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT AS NO PRODUCT HAS BEEN RETURNED. NO CONCLUSION CAN BE DRAWN AT THIS TIME. WE, THEREFORE, CONSIDER THIS REPORT COMPLETE TO THE BEST OF OUR KNOWLEDGE.
Description of Event or Problem · 1
THE CUSTOMER STATED, SHE WAS HOSPITALIZED DUE TO HYPERGLYCEMIA. THE BLOOD GLUCOSE READING WAS NOT REPORTED. TROUBLESHOOTING WAS PERFORMED AND THE PROGRAMMING ON THE INSULIN PUMP WAS CORRECT. THE INSULIN PUMP PASSED THE PRIME TEST. THE CUSTOMER COULD NOT PERFORM THE HIGH PRESSURE TEST BECAUSE SHE DID NOT HAVE A TUBING CLAMP. THE CUSTOMER WAS SHIPPING A TUBING CLAMP AND ADVISED TO CALL BACK TO PERFORM THE HIGH PRESSURE TEST.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | PUMP MMT-522NAS PRDGM INS V2.2 SK EN | INFUSION PUMP | LZG | MEDTRONIC PUERTO RICO OPERATIONS MED-REL | MMT-522NAS |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Hospitalization |