UNICEL DXC 800 PRO SYNCHRON CLINICAL SYSTEMS
Report
- Report Number
- 2050012-2008-00027
- Event Type
- Other
- Date Received
- July 16, 2008
- Date of Event
- June 21, 2008
- Report Date
- July 16, 2008
- Manufacturer
- BECKMAN COULTER, INC.
- Product Code
- JJE
- PMA / PMN Number
- K042291
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- SC, US
- Reporter Occupation
- NO INFORMATION
Narratives
IN 2008, THE AUTOMATED WEEKLY MAINTENANCE PROCEDURE WAS PERFORMED ON THE ION SELECTIVE ELECTRODE (ISE) SYSTEM. THE ISE SYSTEM WAS CALIBRATED AND QC RECOVERED WITHIN THE LAB'S ESTABLISHED RANGES. AT SOME POINT DURING THE ROUTINE OPERATION, THE OPERATOR NOTED THAT THERE WERE FREQUENT DELTA CHECK MESSAGES. QC WAS RUN AND NA QC RECOVERED UP TO 10MMOL/L LOWER THAN THE PREVIOUS RUN. ALL SAMPLES WERE THEN REPEATED ON THE LAB'S SECOND ANALYZER. THE NA RESULTS WERE SIGNIFICANTLY HIGHER AND AMENDED REPORTS WERE ISSUED. THE ISE SYSTEM ON THIS ANALYZER WAS RECALIBRATED: QC WAS WITHIN THE ESTABLISHED RANGES AND NORMAL OPERATION CONTINUED. A FIELD SERVICE ENGINEER (FSE) WORKED WITH THE CUSTOMER BY PHONE AND TOLD THE CUSTOMER THAT A LETTER WILL BE MAILED FROM BCI TO ALL DXC CUSTOMERS REGARDING ISE MAINTENANCE PROCEDURE. UNTIL THEN, THE CUSTOMER WILL PERFORM MORE FREQUENT QC RUNS AND CALIBRATE THE ISE SYSTEM AS NECESSARY. A CLEAR ROOT CAUSE HAS NOT BEEN DETERMINED TO DATE. A MALFUNCTION WILL BE ASSUMED FOR THE PURPOSE OF THIS REPORT.
A CUSTOMER CONTACTED BECKMAN COULTER INC. (BCI) REGARDING ERRONEOUSLY LOW SODIUM (NA) TEST RESULTS PRODUCED BY THE UNICEL DXC 800 PRO SYNCHRON CHEMISTRY ANALYZER FOR SEVERAL PTS. ONLY ONE EXAMPLE OF THE ERRONEOUS RESULTS WAS PROVIDED. THE ORIGINAL RESULT WAS 119 MMOL/L AND THE CORRECTED RESULT WAS 130 MMOL/L. A HOSPITAL CRITICAL EVENT REPORT WAS FILED BECAUSE A PT WAS TREATED BASED ON AN ERRONEOUS NA RESULT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | UNICEL DXC 800 PRO SYNCHRON CLINICAL SYSTEMS | CLINICAL CHEMISTRY ANALYZER | JJE | BECKMAN COULTER, INC. | DXC 800 PRO | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | NA |