OMNIPOD INSULIN PUMP
Report
- Report Number
- 3004464228-2008-00144
- Event Type
- Injury
- Date Received
- July 25, 2008
- Date of Event
- June 21, 2008
- Report Date
- June 26, 2008
- Manufacturer
- INSULET CORPORATION
- Product Code
- LZG
- PMA / PMN Number
- K042792
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Occupation
- NOT APPLICABLE
Narratives
THE PRODUCT HAS NOT BEEN RETURNED SO NO EVALUATION WAS POSSIBLE. THE CUSTOMER FELT RESISTANCE DURING THE FILL PROCESS, WHICH INDICATES A PROBABLE PROBLEM WITH A RETAINER THAT ALLOWED A COMPONENT TO MOVE RESULTING IN DAMAGE WHICH PREVENTED THE PLUNGER FROM ADVANCING. THE USER WOULD HAVE BEEN AWARE OF A PROBLEM DURING THE FILL PROCESS. THERE IS RESISTANCE FELT IN FILLING THE POD WITH INSULIN AND A DISTINCT "CRACKLING" NOISE RESULTING FROM STRIPPING THE THREADS OF THE COMPONENT WHEN OVER-PRESSURISED. THE OMNIPOD USER GUIDE STATES: "WARNING: NEVER USE A POD IF, DURING FILL, YOU DETECT ANY CRACKLING NOISE OR RESISTANCE WHILE DEPRESSING THE PLUNGER OF THE FILL SYRINGE. USING A POD WITH THESE CONDITIONS COULD RESULT IN UNDER-DELIVERY OF INSULIN." THE USER IS ALSO INSTRUCTED IN THE USER GUIDE TO FREQUENTLY MONITOR THEIR BG LEVELS. BY FOLLOWING THESE RECOMMENDATIONS, THE USER WOULD BECOME AWARE OF HIGH BG LEVELS AND COULD USE ANOTHER DEVICE OR BACKUP THERAPY IF REQUIRED.
CUSTOMER CALLED IN REGARDING POD THAT MADE A CLICKING SOUND AND WAS DIFFICULT TO FILL. SHE WAS ADVISED NOT TO USE A POD THAT IS DIFFICULT TO FILL. CUSTOMER STATED THAT SHE WENT TO BED AT MIDNIGHT. AT 5:00 AM, HER BG READ HIGH, SO SHE BOLUSED 6 UNITS WITH THE POD. AT 6:00 AM, SHE BOLUSED 18 UNITS THROUGH SYRINGE. AT 7:30 AM, SHE TOOK THE POD OFF AND WENT TO THE HOSPITAL WHERE SHE WAS PUT ON AN IV UNTIL MONDAY 23RD. NO FURTHER INFORMATION REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | OMNIPOD INSULIN PUMP | PUMP, INFUSION, INSULIN | LZG | INSULET CORPORATION | 11200 | L11689 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Hospitalization |