FDA Adverse Event Malfunction Summary report: N

PROGRAMMING WAND

MDR report key: 1081549 · Received July 25, 2008

Report

Report Number
1644487-2008-01701
Event Type
Malfunction
Date Received
July 25, 2008
Date of Event
January 1, 2008
Report Date
June 26, 2008
Manufacturer
CYBERONICS, INC.
Product Code
LYJ
PMA / PMN Number
P970003
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
WA, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

CONCLUSIONS: DEVICE FAILURE OCCURRED, BUT DID NOT CAUSE OR CONTRIBUTE TO A DEATH OR SERIOUS INJURY.

Description of Event or Problem · 1

IT WAS REPORTED THAT PHYSICIAN'S WAND WAS HAVING PROBLEMS COMMUNICATING WITH THE HANDHELD. A NEW PROGRAMMING WAND WAS RECEIVED AND COMMUNICATION ISSUES HAVE RESOLVED. PRODUCT ANALYSIS WAS PERFORMED ON THE PROGRAMMING WAND. THE SERIAL DATA CABLE, WHICH PRODUCED COMMUNICATION ERRORS, HAD AN INTERMITTENT CONDUCTOR.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 PROGRAMMING WAND LYJ CYBERONICS, INC. 201

Patients

Seq Age Sex Outcome Treatment
1