FDA Adverse Event Injury Summary report: N

STONE CONE NITINOL UROLOGICAL RETRIEVAL COIL

MDR report key: 1081446 · Received July 25, 2008

Report

Report Number
3005099803-2008-01284
Event Type
Injury
Date Received
July 25, 2008
Date of Event
June 25, 2008
Report Date
June 25, 2008
Manufacturer
BOSTON SCIENTIFIC
Product Code
FGO
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
OH, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

THE COMPLAINANT INDICATED THAT THE DEVICE WAS DISPOSED OF AT THE USER FACILITY AND WILL NOT BE RETURNED FOR EVALUATION, THEREFORE, A FAILURE ANALYSIS OF THE COMPLAINT DEVICE COULD NOT BE COMPLETED. A REVIEW OF THE BATCH HISTORY, HISTORICAL TRENDING, AND SIMILAR COMPLAINT TRENDING REVIEW OF THE PRODUCT FAMILY WILL BE CONDUCTED. IF THERE IS ANY FURTHER RELEVANT INFORMATION FROM THAT REVIEW, A SUPPLEMENTAL MEDWATCH WILL BE FILED.

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING A URETEROSCOPY PROCEDURE, A COATING DETACHMENT OCCURRED. THE LESION WAS LOCATED IN THE LEFT KIDNEY. AT AN UNSPECIFIED TIME DURING THE PROCEDURE, SOME OF THE STONE CONE 3FR./7MM RETRIEVAL COIL COATING "CAME OFF IN THE KIDNEY". THE PHYSICIAN WAS ABLE TO REMOVE THE DETACHED PORTION WITH AN UNSPECIFIED GRASPER. A DIFFERENT DEVICE WAS USED TO COMPLETE THE PROCEDURE. THE PT'S STATUS IS REPORTED AS "FINE". IT WAS FURTHER NOTED THAT THE PROCEDURE TOOK AN ADDITIONAL 10 MINUTES.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 STONE CONE NITINOL UROLOGICAL RETRIEVAL COIL FGO - DISLODGER, STONE, FLEXIBLE FGO BOSTON SCIENTIFIC M0063903100 0000605993

Patients

Seq Age Sex Outcome Treatment
1 61 YR Required Intervention