COLLAMER ULTRAVIOLET-ABSORBING POSTERIOR CHAMBER THREE PIECE FOLDABLE INTRAOCULA

MDR report key: 1081249 · Received July 24, 2008

Report

Report Number
2023826-2008-00972
Event Type
Injury
Date Received
July 24, 2008
Date of Event
June 24, 2008
Report Date
June 25, 2008
Manufacturer
STAAR SURGICAL CO.
Product Code
HQL
PMA / PMN Number
P990013
Removal / Correction Number
NA
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
AZ, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

EVAL CODES, RESULTS: VISUAL INSPECTION OF THE RETURNED PRODUCT SHOWED THAT THE OPTIC WAS TORN AND A HAPTIC WAS TORN OFF FROM THE OPTIC AND DEFORMED. THERE WAS A CLEAR SURGICAL RESIDUE ON THE LENS. CONCLUSION: A MULTIFUNCTIONAL TEAM INVESTIGATED COMPLAINTS REGARDING LENS TEARS ASSOCIATED WITH THE CQ CARTRIDGE: THE RESULTANT CORRECTIVE ACTIONS INCLUDED IMPROVEMENTS IN MANUFACTURING, RELEASE TESTING, AND EVAL OF TEST RESULTS. ADDITIONALLY, THE INJECTOR/CARTRIDGE DIRECTIONS FOR USE (DFU) HAVE BEEN MODIFIED TO ADD FURTHER CLARIFICATIONS TO INSTRUCT THE USERS IN THE PROPER DELIVERY TECHNIQUES THAT ARE EFFECTIVE, AND MINIMIZE THE POTENTIAL FOR DAMAGING THE LENS.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE SURGEON INSERTED A CQ2015A THREE PIECE COLLAMER LENS AND A HAPTIC TORE UPON INSERTION. THE INCISION WAS ENLARGED TO REMOVE THE LENS AND SUTURED AFTER ANOTHER LENS WAS IMPLANTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 COLLAMER ULTRAVIOLET-ABSORBING POSTERIOR CHAMBER THREE PIECE FOLDABLE INTRAOCULA INTRAOCULAR LENS HQL STAAR SURGICAL CO. CQ2015A NA

Patients

Seq Age Sex Outcome Treatment
1 Required Intervention STAARVISC II| CARTRIDGE MODEL CQ CARTRIDGE-FP| INJECTOR MODEL MSI-PM