FDA Adverse Event Injury Summary report: N

MBT CEM KEEL TIB TRAY SZ2.5

MDR report key: 1081181 · Received July 23, 2008

Report

Report Number
1818910-2008-02737
Event Type
Injury
Date Received
July 23, 2008
Date of Event
June 23, 2008
Report Date
June 23, 2008
Manufacturer
DEPUY-CORK, A DIVISION OF DEPUY ORTHOPAEDICS
Product Code
NJL
PMA / PMN Number
P830055/S095
Removal / Correction Number
NA
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
NC, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

EXAMINATION WAS NOT POSSIBLE, AS NO PRODUCT WAS RETURNED. A SEARCH OF THE WARSAW AND INTERNATIONAL COMPLAINT DATABASE DID NOT FIND ANY ADDITIONAL REPORTS OF THIS NATURE FOR THE PRODUCT CODE/LOTS PROVIDED SINCE THEIR RESPECTIVE RELEASE FOR DISTRIBUTION. THE INVESTIGATION COULD NOT VERIFY OR IDENTIFY ANY EVIDENCE OF PRODUCT CONTRIBUTION TO THE REPORTED PROBLEM. BASED ON THE INVESTIGATION, THE NEED FOR CORRECTIVE ACTION IS NOT INDICATED. DEPUY CONSIDERS THE INVESTIGATION CLOSED AT THIS TIME. SHOULD THE PRODUCT AND/OR ADDITIONAL INFO BE RECEIVED TO CHANGE THE OUTCOME OF THE PERFORMED INVESTIGATION, THE COMPLAINT WILL BE RE-OPENED.

Description of Event or Problem · 1

PATIENT WAS REVISED TO ADDRESS PAIN.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 MBT CEM KEEL TIB TRAY SZ2.5 87NJL NJL DEPUY-CORK, A DIVISION OF DEPUY ORTHOPAEDICS NA 2257728

Patients

Seq Age Sex Outcome Treatment
1 69 YR Required Intervention