FDA Adverse Event Malfunction Summary report: N

EASYPUMP

MDR report key: 10810498 · Received November 9, 2020

Report

Report Number
9610825-2020-00289
Event Type
Malfunction
Date Received
November 9, 2020
Report Date
January 26, 2021
Manufacturer
B. BRAUN MELSUNGEN AG
Product Code
MEB
UDI-DI
04046964448942
PMA / PMN Number
K081905
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 0

THIS REPORT HAS BEEN IDENTIFIED AS B. BRAUN MEDICAL INC. INTERNAL REPORT NUMBER (B)(4). ONE (1) SAMPLE WAS RECEIVED FOR EVALUATION. PARTICULATE WAS FOUND IN THE SOLUTION. THE PARTICULATE WAS EXTRACTED AND MEASURED 0.297MM2, WHICH IS WITHIN THE PRODUCT SPECIFICATION. THE PARTICULATE WAS SENT FOR FTIR ANALYSIS TO DETERMINE THE COMPOSITION OF THE PARTICULATE. THE RESULTS OF THIS TESTING DID NOT MATCH ANY OF THE KNOWN SUBSTANCES WITHIN THE PRODUCTION PROCESS, AND COULD NOT BE IDENTIFIED AT THIS TIME. THE PARTICULATE MATTER IN THE RETURNED SAMPLE WAS FOUND ABOVE THE DEVICE'S EIS FILTER. THE EIS FILTER CAN FILTER PARTICLES WITH THE SIZE OF 1.2 M. THE PARTICULATE FOUND IN THE RETURNED SAMPLE WAS >1.2 M, AND WOULD BE FILTERED OUT BY THE EIS FILTER. BASED ON THE RESULTS OF THE INVESTIGATION, NO CONCLUSIONS CAN BE MADE REGARDING THE CAUSE OF THE REPORTED EVENT.

Additional Manufacturer Narrative · 1

THIS REPORT HAS BEEN IDENTIFIED AS B. BRAUN MEDICAL INC. INTERNAL REPORT NUMBER 400491825. THE INVESTIGATION INTO THIS REPORTED EVENT IS ONGOING. ADDITIONAL ATTEMPTS TO RECEIVE THE DEVICE INVOLVED IN THE REPORTED EVENT ARE BEING MADE. A FOLLOW-UP REPORT WILL BE SUBMITTED WHEN THE RESULTS OF THE INVESTIGATION ARE AVAILABLE.

Description of Event or Problem · 1

AS REPORTED BY USER FACILITY: BLACK PARTICLE IN PUMP. ISSUE DISCOVERED IN THE PHARMACY DURING COMPOUNDING PROCESS.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1274809 EASYPUMP PUMP, INFUSION, ELASTOMERIC MEB B. BRAUN MELSUNGEN AG 4540040-02 19N18GE672 04046964448942

Patients

Seq Age Sex Outcome Treatment
1