FDA Adverse Event Injury Summary report: N

TALENT ABDOMINAL STENT GRAFT SYSTEM

MDR report key: 1080770 · Received July 23, 2008

Report

Report Number
2953200-2008-00593
Event Type
Injury
Date Received
July 23, 2008
Date of Event
June 25, 2008
Report Date
June 25, 2008
Manufacturer
MEDTRONIC CARDIOVASCULAR GALWAY
Product Code
MIH
PMA / PMN Number
P070027
Removal / Correction Number
NA
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
TX, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

REQUIRED SECONDARY INTERVENTION.

Description of Event or Problem · 1

A TALENT ABDOMINAL STENT GRAFT SYSTEM WAS IMPLANTED IN A PATIENT FOR THE ENDOVASCULAR TREATMENT OF AN ABDOMINAL AORTIC ANEURYSM. ANEURYSM MORPHOLOGY WAS NOT REPORTED AND THE AORTIC NECK WAS 23 - 25 MM IN DIAMETER AND IT HAD AN ANGULATION OF 30 - 40 DEGREES. IT WAS REPORTED THAT AFTER THE BIFURCATED STENT GRAFT WAS IMPLANTED THERE WAS A TYPE 1 ENDOLEAK, DUE TO INFOLDING OF THE STENT GRAFT. MODELING WITH RELIANT BALLOON WAS PERFORMED AND THAT IMPROVED THE ENDOLEAK TO SOME EXTENT, BUT DID NOT COMPLETELY RESOLVE IT. THE ENDOLEAK WAS RESOLVED BY PLACEMENT OF AN ANEURX AORTIC CUFF. NO ADDITIONAL CLINICAL SEQUELAE WERE REPORTED AND THE PATIENT IS FINE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 TALENT ABDOMINAL STENT GRAFT SYSTEM MIH MEDTRONIC CARDIOVASCULAR GALWAY NA V00114814

Patients

Seq Age Sex Outcome Treatment
1 UNK Required Intervention