FDA Adverse Event Malfunction Summary report: N

CHANNEL ROUND 15FR DRAIN WITH TROCAR

MDR report key: 10807679 · Received November 9, 2020

Report

Report Number
8030107-2020-00039
Event Type
Malfunction
Date Received
November 9, 2020
Report Date
November 9, 2020
Manufacturer
DEGANIA SILICONE LTD
Product Code
OTK
PMA / PMN Number
EXEMPT
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
JA
Reporter Occupation
RISK MANAGER

Narratives

Additional Manufacturer Narrative · 1

SINCE THE IN-PROCESS REPORTS AND TEST OF RETAIN SAMPLE FROM THE SAME BATCH SHOWED OK RESULTS, WE CONCLUDE THAT THE DRAIN POSSIBLY COULD TEAR FOLLOWING SOME DAMAGE IN USE.

Description of Event or Problem · 1

BROKEN/CRACKED/FRACTURED DURING USE:FRACTURE AT THE DRAIN TUBE JOINT DURING USE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1272992 CHANNEL ROUND 15FR DRAIN WITH TROCAR SILICONE FLUTED ROUND DRAIN 15FR OTK DEGANIA SILICONE LTD 15FR P1760471

Patients

Seq Age Sex Outcome Treatment
1