FDA Adverse Event Other Summary report: N

3003768277-2008-00018

MDR report key: 1080711 · Received July 9, 2008

Report

Report Number
3003768277-2008-00018
Event Type
Other
Date Received
July 9, 2008
Product Code
LNH
PMA / PMN Number
k052013
Report Source
Manufacturer report

Narratives

Additional Manufacturer Narrative · 1

(CONCLUSIONS)- CODE (OTHER)- THE RF BURN WAS MOST LIKELY CAUSE BY UNWANTED RF INTERFERENCE. THE INTERNAL SHIELDING OF THE COIL MAY BE ALSO DAMAGED, DEPENDING OF THE POSITION OF THE COIL AND THE PT THIS COULD LEAD TO HEATING OF THE COIL ELEMENTS CAUSING THE RF BURN. THIS IS NO PROBLEM IN CASE ISOLATING MATERIAL IS USED, BUT MAY LEAD TO BURNS IN CASE THE COIL IS USED WITHOUT PROTECTION. THESE KIND OF INCIDENTS CAN BE PREVENTED AS MUCH AS POSSIBLE, BY FOLLOWING THE INSTRUCTIONS FOR USE.