FDA Adverse Event
Other
Summary report: N
ACHIEVA 1.5T NOVA
MDR report key: 1080710
·
Received July 9, 2008
Report
- Report Number
- 1217116-2008-00045
- Event Type
- Other
- Date Received
- July 9, 2008
- Date of Event
- February 12, 2008
- Report Date
- March 20, 2008
- Manufacturer
- PHILIPS MEDICAL SYSTEMS
- Product Code
- LNH
- Adverse Event
- Yes
- Report Source
- Distributor report
- Reporter Location
- NL
- Reporter Occupation
- OTHER
Narratives
Description of Event or Problem · 1
THIS REPORT IS BEING FILED UPON NOTIFICATION FROM OUR MR MFR, OF AN EVENT THAT OCCURRED IN ANOTHER COUNTRY. THE PT WAS SCANNED WITH THE FEET FIRST INTO THE MAGNET. THE PT WAS AN AUTISTIC PERSON AND WAS ON GENERAL ANESTHETIC DURING THE EXAMINATION. THEREFORE, ECG-LEADS, OXYGEN SATURATION, AND A TUBE FOR ANESTHESIA WERE ALSO CONNECTED TO THE PT. AFTER THE SCAN A SECOND DEGREE BURN WITH BLISTER (5CM) APPEARED ON THE RIGHT ARM, BETWEEN THE WRIST AND ELBOW WHERE THE ARM WAS TOUCHING THE CABLE OF THE RF-COIL.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | ACHIEVA 1.5T NOVA | LNH ( MAGNETIC RESONANCE DIAGNOSTIC DEVICE) | LNH | PHILIPS MEDICAL SYSTEMS | 781196 | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 29 YR | Other |