FDA Adverse Event Injury Summary report: N

PHOENIX

MDR report key: 1080685 · Received July 11, 2008

Report

Report Number
2087532-2008-00062
Event Type
Injury
Date Received
July 11, 2008
Date of Event
May 15, 2008
Report Date
June 4, 2008
Manufacturer
GAMBRO DASCO
Product Code
FII
Adverse Event
Yes
Product Problem
Yes
Report Source
Distributor report
Reporter Location
MO, US
Reporter Occupation
NURSE

Narratives

Description of Event or Problem · 1

GAMBRO RECEIVED A REPORT FROM A CUSTOMER WHO COMPLAINED THAT THREE NURSES SUSTAINED BURNS WHEN CONCENTRATE TUBING CAME OFF OF THE BLUE MALE CONCENTRATE CONNECTOR ASSEMBLY INSERTED INTO THE LOWER FRONT PANEL ON A PHOENIX MACHINE. THE EVENTS OCCURRED ON THREE DIFFERENT MACHINES WHICH WERE ALL EQUIPPED WITH THE SOFTWARE VERSION 3.34. (FOR THE OTHER EVENTS REFER TO MDR 9616240-2008-00036 AND MFR 9616240-2008-00037). HOT WATER WAS SPRAYED ONTO THE RIGHT LOWER FOREARM OF THE FACILITY'S NURSE WHILE THE DEVICE WAS UNDERGOING HEAT DISINFECTION. AT THAT MOMENT, THE NURSE WAS WIPING DOWN THE MACHINE WITH A DILUTE BLEACH SOLUTION. THE NURSE WAS DIAGNOSED WITH FIRST- DEGREE BURNS AND WAS TREATED AT THE EMERGENCY DEPT AT THE HOSPITAL. THE MACHINE WAS INSPECTED BY THE CLINIC'S BIOMEDICAL TECHNICAN WHO FOUND THAT THE CONCENTRATE TUBING AND THE TUBING CLIP HAD COME LOOSE FROM THE MALE CONNECTOR ASSEMBLY. HE REPLACED THE TUBING AND THE CLIP BACK ONTO THE MALE CONNECTOR ASSEMBLY. THE MACHINE WAS RETURNED TO SERVICE WITHOUT ANY FURTHER TUBE DISCONNECTION EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 PHOENIX HEMODIALYSIS EQUIPMENT FII GAMBRO DASCO PHOENIX NA

Patients

Seq Age Sex Outcome Treatment
1 UNKNOWN Required Intervention