FDA Adverse Event Other Summary report: N

DA VINCI S SURGICAL SYSTEM

MDR report key: 1080670 · Received July 21, 2008

Report

Report Number
2955842-2008-01153
Event Type
Other
Date Received
July 21, 2008
Date of Event
June 23, 2008
Report Date
July 18, 2008
Manufacturer
INTUITIVE SURGICAL, INC.
Product Code
NAY
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NE, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

THE INVESTIGATION CONDUCTED BY AN ISI REPRESENTATIVE CONFIRMED THAT THE ENDOSCOPIC CAMERA MANIPULATOR (ECM) CANNULA MOUNT WAS BROKEN. THE ECM CANNULA MOUNT HOLDS THE CANNULA SECURELY IN PLACE WHILE THE CAMERA IS DELIVERED THROUGH THE CANNULA. THE CANNULA PROVIDES THE PORT THROUGH THE BODY WALL OF THE PT, THROUGH WHICH THE ENDOSCOPE IS INSERTED. THE SYSTEM WAS REPAIRED BY REPLACING THE AFFECTED ECM CANNULA CAMERA MOUNT. AS OF 2007, THERE HAVE BEEN NO REPORTED RECURRENCES OF THE ISSUE AT THIS HOSPITAL.

Description of Event or Problem · 1

IT WAS REPORTED THAT WHILE SETTING UP THE DA VINCI S SURGICAL SYSTEM FOR A HYSTERECTOMY PROCEDURE, THE ENDOSCOPIC CAMERA MANIPULATOR (ECM) CANNULA CAMERA MOUNT BROKE OFF. THE PT WAS UNDER ANESTHESIA FOR APPROXIMATELY 30 MINUTES AND THE PORT INCISIONS HAD BEEN CREATED WHEN THE SITE DECIDED TO CONVERT TO LAPAROSCOPIC SURGICAL TECHNIQUES TO FINISH THE PLANNED SURGERY. NO PT HARM WAS REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 DA VINCI S SURGICAL SYSTEM ENDOSCOPIC INSTRUMENT CONTROL SYSTEM NAY INTUITIVE SURGICAL, INC. IS2000 A5.1P4

Patients

Seq Age Sex Outcome Treatment
1 Other ACCESSORIES| INSTRUMENTS| DA VINCI S SURGICAL SYSTEM