FDA Adverse Event Malfunction Summary report: N

LAP-BAND ADJUSTABLE GASTRIC BANDING SYSTEM 10.0CM

MDR report key: 1080661 · Received July 21, 2008

Report

Report Number
2024601-2008-00382
Event Type
Malfunction
Date Received
July 21, 2008
Date of Event
June 12, 2008
Report Date
June 23, 2008
Manufacturer
COSTA RICA
Product Code
LTI
PMA / PMN Number
P000008
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
AL, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

TAPER II. VISUAL EXAMINATION OF THE RETURNED DEVICE DETERMINED THE ACCESS PORT TUBING CONNECTOR TO BE A TAPER II. ANALYSIS OF THE DEVICE NOTED NON PENETRATING NICKS TO THE PORT AND DAMAGE FROM A SHARP INSTRUMENT TO THE BAND TUBING (THIS IS THE PORTION OF TUBING LOCATED BETWEEN THE STAINLESS STEEL CONNECTOR AND THE BAND, NOT BETWEEN THE STAINLESS STEEL CONNECTOR AND THE PORT). THERE WAS NO INDICATION OF WEAR RELATED DAMAGE TO THE PORTION OF THE LAP-BAND SYSTEM RETURNED. THE LAB WAS UNABLE TO DUPLICATE OR CONFIRM THE REPORTED EVENT. THERE IS NO OTHER INFORMATION AVAILABLE AT THIS TIME FROM THE SURGEON TO DETERMINE THE CAUSE OF THE ALLEGED EVENT. DEVICE LABELING ADDRESSES THE POSSIBLE OUTCOME OF LEAKAGE AS FOLLOWS: "DEFLATION OF THE BAND MAY OCCUR DUE TO LEAKAGE FROM THE BAND, THE PORT OF THE CONNECTOR TUBING."

Description of Event or Problem · 1

REPORTED AS, THE DOCTOR TRIED TO INJECT THE PORT WITH FLUID, AND IT WOULD NOT HOLD FLUID.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 LAP-BAND ADJUSTABLE GASTRIC BANDING SYSTEM 10.0CM DEVICE FOR TREATMENT OF MORBID OBESITY LTI COSTA RICA NA 1396074

Patients

Seq Age Sex Outcome Treatment
1 40 YR