LAP-BAND ADJUSTABLE GASTRIC BANDING SYSTEM 10.0CM
Report
- Report Number
- 2024601-2008-00382
- Event Type
- Malfunction
- Date Received
- July 21, 2008
- Date of Event
- June 12, 2008
- Report Date
- June 23, 2008
- Manufacturer
- COSTA RICA
- Product Code
- LTI
- PMA / PMN Number
- P000008
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- AL, US
- Reporter Occupation
- OTHER
Narratives
TAPER II. VISUAL EXAMINATION OF THE RETURNED DEVICE DETERMINED THE ACCESS PORT TUBING CONNECTOR TO BE A TAPER II. ANALYSIS OF THE DEVICE NOTED NON PENETRATING NICKS TO THE PORT AND DAMAGE FROM A SHARP INSTRUMENT TO THE BAND TUBING (THIS IS THE PORTION OF TUBING LOCATED BETWEEN THE STAINLESS STEEL CONNECTOR AND THE BAND, NOT BETWEEN THE STAINLESS STEEL CONNECTOR AND THE PORT). THERE WAS NO INDICATION OF WEAR RELATED DAMAGE TO THE PORTION OF THE LAP-BAND SYSTEM RETURNED. THE LAB WAS UNABLE TO DUPLICATE OR CONFIRM THE REPORTED EVENT. THERE IS NO OTHER INFORMATION AVAILABLE AT THIS TIME FROM THE SURGEON TO DETERMINE THE CAUSE OF THE ALLEGED EVENT. DEVICE LABELING ADDRESSES THE POSSIBLE OUTCOME OF LEAKAGE AS FOLLOWS: "DEFLATION OF THE BAND MAY OCCUR DUE TO LEAKAGE FROM THE BAND, THE PORT OF THE CONNECTOR TUBING."
REPORTED AS, THE DOCTOR TRIED TO INJECT THE PORT WITH FLUID, AND IT WOULD NOT HOLD FLUID.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | LAP-BAND ADJUSTABLE GASTRIC BANDING SYSTEM 10.0CM | DEVICE FOR TREATMENT OF MORBID OBESITY | LTI | COSTA RICA | NA | 1396074 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 40 YR |