ULTRACISION HARMONIC SCALPEL HAND PIECE
MDR report key: 1080358
·
Received July 24, 2008
Report
- Report Number
- 3005075853-2008-00371
- Event Type
- Malfunction
- Date Received
- July 24, 2008
- Date of Event
- May 21, 2008
- Report Date
- May 21, 2008
- Manufacturer
- ETHICON ENDO-SURGERY, LLC
- Product Code
- LFL
- PMA / PMN Number
- K002906
- Removal / Correction Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CA, US
- Reporter Occupation
- PHYSICIAN
Narratives
Additional Manufacturer Narrative · 1
DATE SENT: 07/10/2008. EVAL SUMMARY: THE HAND PIECE WAS RETURNED WITH THE LOOSE MOUNT AND THE NOSE CONE CRACKED. IT WAS TESTED ON A GENERATOR AND GAVE AN ERROR CODE 3. FOR FURTHER ANALYSIS, THE HAND PIECE WAS DISASSEMBLED, A TORN ACOUSTIC ISOLATOR AND MOISTURE INGRESS WERE FOUND IN THE INSTRUMENT. COMPLAINT INFO IS TRENDED ON A REGULAR BASIS TO DETERMINE IF FURTHER INVESTIGATION IS WARRANTED. THE BATCH HISTORY RECORDS WERE REVIEWED WITH NO ANOMALIES NOTED DURING THE MFG PROCESS.
Description of Event or Problem · 1
IT WAS REPORTED THAT DURING A LIVER RESECTION PROCEDURE, THE DEVICE WENT INTO STANDBY AND THERE IS A CRACK IN THE HOUSING. ANOTHER DEVICE WAS USED TO COMPLETE THE PROCEDURE. THERE WAS NO PT CONSEQUENCE.