FDA Adverse Event Malfunction Summary report: N

ULTRACISION HARMONIC SCALPEL HAND PIECE

MDR report key: 1080358 · Received July 24, 2008

Report

Report Number
3005075853-2008-00371
Event Type
Malfunction
Date Received
July 24, 2008
Date of Event
May 21, 2008
Report Date
May 21, 2008
Manufacturer
ETHICON ENDO-SURGERY, LLC
Product Code
LFL
PMA / PMN Number
K002906
Removal / Correction Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

DATE SENT: 07/10/2008. EVAL SUMMARY: THE HAND PIECE WAS RETURNED WITH THE LOOSE MOUNT AND THE NOSE CONE CRACKED. IT WAS TESTED ON A GENERATOR AND GAVE AN ERROR CODE 3. FOR FURTHER ANALYSIS, THE HAND PIECE WAS DISASSEMBLED, A TORN ACOUSTIC ISOLATOR AND MOISTURE INGRESS WERE FOUND IN THE INSTRUMENT. COMPLAINT INFO IS TRENDED ON A REGULAR BASIS TO DETERMINE IF FURTHER INVESTIGATION IS WARRANTED. THE BATCH HISTORY RECORDS WERE REVIEWED WITH NO ANOMALIES NOTED DURING THE MFG PROCESS.

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING A LIVER RESECTION PROCEDURE, THE DEVICE WENT INTO STANDBY AND THERE IS A CRACK IN THE HOUSING. ANOTHER DEVICE WAS USED TO COMPLETE THE PROCEDURE. THERE WAS NO PT CONSEQUENCE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 ULTRACISION HARMONIC SCALPEL HAND PIECE NONE LFL ETHICON ENDO-SURGERY, LLC NA NA

Patients

Seq Age Sex Outcome Treatment
1 GENERATOR