FDA Adverse Event Malfunction Summary report: N

LIGAMAX-5MM ENDO CLIP APPLIER

MDR report key: 1080351 · Received July 24, 2008

Report

Report Number
3005075853-2008-00368
Event Type
Malfunction
Date Received
July 24, 2008
Date of Event
June 5, 2008
Report Date
June 16, 2008
Manufacturer
ETHICON ENDO-SURGERY, LLC
Product Code
FZP
PMA / PMN Number
K050344
Removal / Correction Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
JA
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

MALFORMED CLIP. EVAL SUMMARY: THE ANALYSIS RESULTS FOUND THAT THE EL5ML DEVICE WAS RETURNED IN GOOD VISUAL CONDITION. THE INSTRUMENT WAS CYCLED, FED AND FORMED 5 CLIP CONFORMING AND 1 MALFORMED CLIP DUE TO BYPASS ISSUES. WE HAVE DOCUMENTED THE CIRCUMSTANCES AS THEY WERE REPORTED TO US. IN ADDITION, COMPLAINT INFO IS TRENDED ON A REGULAR BASIS TO DETERMINE IF FURTHER INVESTIGATION IS WARRANTED. A BATCH HISTORY REVIEW WAS PERFORMED AND NO ANOMALIES WERE FOUND.

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING A CHOLECYSTECTOMY PROCEDURE THE CLIP WAS NOT FED INTO THE JAWS PROPERLY AND IT COULD NOT BE FIRED. ANOTHER DEVICE WAS USED TO COMPLETE THE CASE WITH NO PT CONSEQUENCE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 LIGAMAX-5MM ENDO CLIP APPLIER NONE FZP ETHICON ENDO-SURGERY, LLC NA E4KT77

Patients

Seq Age Sex Outcome Treatment
1