FDA Adverse Event
Malfunction
Summary report: N
LIGAMAX-5MM ENDO CLIP APPLIER
MDR report key: 1080351
·
Received July 24, 2008
Report
- Report Number
- 3005075853-2008-00368
- Event Type
- Malfunction
- Date Received
- July 24, 2008
- Date of Event
- June 5, 2008
- Report Date
- June 16, 2008
- Manufacturer
- ETHICON ENDO-SURGERY, LLC
- Product Code
- FZP
- PMA / PMN Number
- K050344
- Removal / Correction Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- JA
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
MALFORMED CLIP. EVAL SUMMARY: THE ANALYSIS RESULTS FOUND THAT THE EL5ML DEVICE WAS RETURNED IN GOOD VISUAL CONDITION. THE INSTRUMENT WAS CYCLED, FED AND FORMED 5 CLIP CONFORMING AND 1 MALFORMED CLIP DUE TO BYPASS ISSUES. WE HAVE DOCUMENTED THE CIRCUMSTANCES AS THEY WERE REPORTED TO US. IN ADDITION, COMPLAINT INFO IS TRENDED ON A REGULAR BASIS TO DETERMINE IF FURTHER INVESTIGATION IS WARRANTED. A BATCH HISTORY REVIEW WAS PERFORMED AND NO ANOMALIES WERE FOUND.
Description of Event or Problem · 1
IT WAS REPORTED THAT DURING A CHOLECYSTECTOMY PROCEDURE THE CLIP WAS NOT FED INTO THE JAWS PROPERLY AND IT COULD NOT BE FIRED. ANOTHER DEVICE WAS USED TO COMPLETE THE CASE WITH NO PT CONSEQUENCE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | LIGAMAX-5MM ENDO CLIP APPLIER | NONE | FZP | ETHICON ENDO-SURGERY, LLC | NA | E4KT77 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |