FDA Adverse Event
Malfunction
Summary report: N
ULTRACISION HARMONIC SCALPEL HAND PIECE
MDR report key: 1080331
·
Received July 24, 2008
Report
- Report Number
- 3005075853-2008-00294
- Event Type
- Malfunction
- Date Received
- July 24, 2008
- Date of Event
- April 18, 2008
- Report Date
- April 18, 2008
- Manufacturer
- ETHICON ENDO-SURGERY, LLC.
- Product Code
- LFL
- PMA / PMN Number
- K002906
- Removal / Correction Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CA, US
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
DATE SENT: 07/08/2008. THE HAND PIECE WAS RETURNED WITH A LOOSE MOUNT AND THE NOSE CONE CRACKED. THE ANALYSIS WAS PERFORMED AND WAS FOUND THAT THE HAND PIECE NO LONGER MEETS THE SPECIFICATIONS FOR CONTINUITY. IT WAS TESTED ON A GENERATOR AND NO ERROR CODE WAS NOTED. THE HAND PIECE WAS DISASSEMBLED AND FOUND A TORN ACOUSTIC ISOLATOR AND MOISTURE INGRESS INSIDE OF IT. COMPLAINT INFO IS TRENDED ON A REGUALR BASIS TO DETERMINE IF FURTHER INVESTIGATION IS WARRANTED. THE BATCH HISTORY RECORDS WERE REVIEWED WITH NO ANOMALIES NOTED DURING THE MANUFACTURING PROCESS.
Description of Event or Problem · 1
IT WAS REPORTED THAT DURING A LAP GASTRIC BYPASS PROCEDURE, THE HANDPIECE HOUSING WAS CRACKED. ANOTHER DISPOSABLE HAD TO BE PULLED. THERE WAS NO PT CONSEQUENCE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | ULTRACISION HARMONIC SCALPEL HAND PIECE | NONE | LFL | ETHICON ENDO-SURGERY, LLC. | NA | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | GENERATOR |