FDA Adverse Event Other Summary report: N

DIMENSION CLINICAL CHEMISTRY SYSTEM

MDR report key: 1080219 · Received July 18, 2008

Report

Report Number
2517506-2008-00069
Event Type
Other
Date Received
July 18, 2008
Date of Event
June 20, 2008
Report Date
June 20, 2008
Manufacturer
SIEMENS HEALTHCARE DIAGNOSTICS INC.,
Product Code
JJE
PMA / PMN Number
K010313
Removal / Correction Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NY, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

ANALYSIS OF THE INSTRUMENT AND INSTRUMENT DATA INDICATE THAT THE CAUSE FOR THE FALSELY DEPRESSED TROPONIN I RESULT WAS SAMPLE INTEGRITY. THE IFU FOR DIMENSION CTNI FLEX REAGENT CARTRIDGE CONTAINS THE FOLLOWING INFORMATION: "SPECIMENS SHOULD BE FREE TO PARTICULATE MATTER TO PREVENT APPEARANCE OF FIBRIN IN SERUM SAMPLES, COMPLETE CLOT FORMATION SHOULD TAKE PLACE BEFORE CENTRIFUGATION." THE INSTRUMENT IS PERFORMING WITHIN SPECIFICATIONS. NO FURTHER EVALUATION OF THE DEVICE IS REQUIRED.

Description of Event or Problem · 1

A FALSELY ELEVATED TROPONIN I RESULT WAS OBTAINED ON A PATIENT SAMPLE. THE RESULT WAS NOT REPORTED TO THE PHYSICIAN. THE SAMPLE WAS REPEATED AND A NEGATIVE RESULT WAS OBTAINED. PATIENT TREATMENT WAS NOT ALTERED OR PRESCRIBED. THERE WAS NO REPORT OF ADVERSE HEALTH CONSEQUENCES AS A RESULT OF THE FALSELY ELEVATED TROPONIN I RESULT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 DIMENSION CLINICAL CHEMISTRY SYSTEM CTNI FLEX REAGENT CARTRIDGE JJE SIEMENS HEALTHCARE DIAGNOSTICS INC., NA GB9127

Patients

Seq Age Sex Outcome Treatment
1