FDA Adverse Event Malfunction Summary report: N

M SERIES

MDR report key: 1080154 · Received July 17, 2008

Report

Report Number
1220908-2008-01533
Event Type
Malfunction
Date Received
July 17, 2008
Report Date
June 30, 2008
Manufacturer
ZOLL MEDICAL CORPORATION
Product Code
MKJ
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE REPORTED MALFUNCTION WAS OBSERVED AND ATTRIBUTED TO A MISALIGNED BIPHASIC FLEX CABLE. ANALYSIS OF FAILURES OF THIS TYPE HAS NOT IDENTIFIED AN INCREASE IN TREND.

Description of Event or Problem · 1

COMPLAINANT ALLEGED THAT DURING A ROUTINE SHIFT CHECK BY A CLINICIAN, THE DEVICE DISPLAYED A "UNIT FAILED" MESSAGE. COMPLAINANT INDICATED THAT THERE WAS NO PATIENT INVOLVEMENT IN THE REPORTED MALFUNCTION.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 M SERIES DEFIBRILLATOR/PACEMAKER MKJ ZOLL MEDICAL CORPORATION MSERIES BI-PHASIC NA

Patients

Seq Age Sex Outcome Treatment
1 NA