FDA Adverse Event Malfunction Summary report: N

BD RAPID DETECTION OF SARS-COV-2 VERITOR

MDR report key: 10801314 · Received November 6, 2020

Report

Report Number
1119779-2020-00922
Event Type
Malfunction
Date Received
November 6, 2020
Date of Event
October 19, 2020
Report Date
May 19, 2021
Manufacturer
BECTON, DICKINSON & CO. (SPARKS)
Product Code
QKP
PMA / PMN Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
SD, US
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 0

EUA#: (B)(4). H6: INVESTIGATION SUMMARY BD HAS RECEIVED SEVERAL CUSTOMER COMPLAINTS FOR FALSE POSITIVE RESULTS, WHEN USING BD SARS-COV-2 REAGENTS FOR BD VERITOR¿ SYSTEM. THE CURRENT INVESTIGATION CONCERNS MULTIPLE LOTS OF BD SARS-COV-2 REAGENTS FOR BD VERITOR¿ SYSTEM. BD TAKES A SYSTEMATIC APPROACH TO INVESTIGATING FALSE POSITIVE COMPLAINTS THAT ARE RECEIVED. THIS APPROACH INVOLVES REVIEW OF MANUFACTURING BATCH HISTORY RECORDS, TESTING OF RETENTION SAMPLES, AND TESTING OF CUSTOMER RETURNED SAMPLES IF APPLICABLE. DUE TO UNKNOWN LOT# A DHR COULD NOT BE PERFORMED. THE INVESTIGATION DID NOT FIND A ROOT CAUSE FOR THE FALSE POSITIVE RESULTS THAT WERE OBSERVED. IT IS RECOMMENDED THAT EACH CUSTOMER REVIEW THEIR WORKFLOW CAREFULLY TO ENSURE THAT THE PACKAGE INSERT IS BEING FOLLOWED AS WRITTEN. THE ROOT CAUSE IS UNDER INVESTIGATION AND WILL BE DOCUMENTED IN OUR QUALITY SYSTEM. BD CANNOT CONFIRM THE COMPLAINT BASED ON THE INVESTIGATION THAT WAS PERFORMED. A CORRECTIVE AND PREVENTIVE ACTION (CAPA#1878253) IS ALREADY INITIATED TO INVESTIGATE THE ROOT CAUSE AND SOME MITIGATION ACTIONS ARE ALREADY BEING ADDRESSED. SEE H10.

Description of Event or Problem · 0

IT WAS REPORTED WHILE TESTING FOR SARS-COV-2 A POTENTIAL FALSE POSITIVE RESULT WAS OBTAINED. A NEW SAMPLE WAS RECOLLECTED AND REPEATED ON THE VERITOR. UPON REPEAT THE RESULT WAS NEGATIVE. CONFIRMATION TESTING WAS PERFORMED VIA PCR AND THE RESULT WAS POSITIVE. THE CUSTOMER STATED THE STAFF MEMBER TESTED WAS ASYMPTOMATIC. THIS TEST IS NOT INTENDED FOR USE ON ASYMPTOMATIC PATIENTS AND WAS THEREFORE USED OFF LABEL. THE RESULT WAS REPORTED OUT AND THE STAFF MEMBER WAS SENT HOME TO SELF QUARANTINE. EUA # (B)(4).

Additional Manufacturer Narrative · 1

EUA #(B)(4). MEDICAL DEVICE LOT #: 0223038 WAS REPORTED, HOWEVER, THIS IS NOT A LOT# MANUFACTURED FOR THIS PRODUCT. MEDICAL DEVICE EXPIRATION DATE: UNKNOWN. A DEVICE EVALUATION IS ANTICIPATED, BUT HAS NOT YET BEGUN. UPON COMPLETION OF THE INVESTIGATION, A SUPPLEMENTAL REPORT WILL BE FILED. DEVICE MANUFACTURE DATE: UNKNOWN. (B)(4).

Description of Event or Problem · 1

IT WAS REPORTED WHILE TESTING FOR SARS-COV-2 A POTENTIAL FALSE POSITIVE RESULT WAS OBTAINED. A NEW SAMPLE WAS RECOLLECTED AND REPEATED ON THE VERITOR. UPON REPEAT THE RESULT WAS NEGATIVE. CONFIRMATION TESTING WAS PERFORMED VIA PCR AND THE RESULT WAS POSITIVE. THE CUSTOMER STATED THE STAFF MEMBER TESTED WAS ASYMPTOMATIC. THIS TEST IS NOT INTENDED FOR USE ON ASYMPTOMATIC PATIENTS AND WAS THEREFORE USED OFF LABEL. THE RESULT WAS REPORTED OUT AND THE STAFF MEMBER WAS SENT HOME TO SELF QUARANTINE. EUA # (B)(4).

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1263991 BD RAPID DETECTION OF SARS-COV-2 VERITOR CORONAVIRUS ANTIGEN DETECTION SYSTEM QKP BECTON, DICKINSON & CO. (SPARKS) SEE H.10

Patients

Seq Age Sex Outcome Treatment
1