DPS DRUCKMESS SET
Report
- Report Number
- 1526863-2008-00024
- Event Type
- Malfunction
- Date Received
- July 17, 2008
- Date of Event
- February 28, 2008
- Report Date
- June 17, 2008
- Manufacturer
- SMITHS MEDICAL ASD, INC.
- Product Code
- DRS
- PMA / PMN Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- UK
- Reporter Occupation
- NOT APPLICABLE
Narratives
THE SUBASSEMBLY IN QUESTION IS MFG BY SMITHS MEDICAL. THIS ASSEMBLY IS INCORPORATED INTO THE FINISHED PROD BY SMITHS MEDICAL INTL IN ANOTHER COUNTRY. THE FINISHED PROD IS FURTHER ASSEMBLED, PACKAGED AND STERILIZED BY SMITHS MEDICAL INTL. TWO SUBASSEMBLIES WERE REC'D. BOTH THE MALE LUER LOCK (MLL) AND FEMALE LURE LOCK (FLL) WERE MISSING FROM THE TUBING ON ONE SAMPLE. THE SECOND UNIT WAS MISSING THE FLL. EXAMINATION OF THE DISCONNECTION AREAS OF BOTH SAMPLES SHOWED THE TUBING HAD BEEN FULLY INSERTED AND ADHESIVE HAD BEEN APPLIED. THE TUBING DIAMETER WAS WITHIN SPEC. THERE WERE NO MFG ISSUES NOTED THAT WOULD HAVE CONTRIBUTED TO THE DISCONNECTIONS. THE TUBING ASSEMBLY IN QUESTION WAS MFG ON AN AUTOMATIC TUBING ASSEMBLY MACHINE. THE PROD IS 100% VISUALLY INSPECTED BY THE OPERATOR FOR PROPER ASSEMBLY AND SUFFICIENT SOLVENT. AS PART OF THE INSPECTION PROCESS ANY UNITS REMOVED DURING THE 100% WERE REWORKED BY THE OPERATOR. AS A PRECAUTION IN MID 2007, THIS PRACTICE HAS BEEN DISCONTINUED. ANY UNITS REMOVED DURING THE 100% INSPECTION ARE BEING DISCARDED. SMITHS WAS ABLE TO CONFIRM THE ISSUE; HOWEVER, NO ROOT CAUSE COULD BE DETERMINED. NO ADD'L ACTION IS NECESSARY AT THIS TIME. SMITHS CONSIDERS THIS MDR CLOSED WITH THIS REPORT.
ANOTHER COUNTRY, REPORTED TO SMITHS MEDICAL INTL THAT THE ARTERIAL PRESSURE LINE DETACHED FROM THE LUER LOCK ON THE STOPCOCK. THIS WAS FOUND AFTER 48 HRS IN USE. THERE WAS NO PT INJURY OR TREATMENT REPORTED WITH THIS EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | DPS DRUCKMESS SET | PRESSURE MONITORING SET | DRS | SMITHS MEDICAL ASD, INC. | NA | 1201439 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | UNK |