FDA Adverse Event Malfunction Summary report: N

ULTRACISION HARMONIC SCALPEL CURVED SHEARS (LCS) PISTOL GRIP 36CM - 15 MM ACTIVE

MDR report key: 1080018 · Received July 17, 2008

Report

Report Number
3005075853-2008-00572
Event Type
Malfunction
Date Received
July 17, 2008
Date of Event
June 24, 2008
Report Date
June 24, 2008
Manufacturer
ETHICON ENDO-SURGERY, LLC.
Product Code
GEI
PMA / PMN Number
K002981
Removal / Correction Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
KY, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE ANALYSIS RESULTS FOUND THAT THE DEVICE WAS RETURNED WITH THE TISSUE PAD MISSING. AS THE TISSUE PAD WAS NOT RETURNED, FURTHER EVALUATION COULD NOT BE PERFORMED. EACH DEVICE IS VISUALLY INSPECTED AND FUNCTIONALLY TESTED PRIOR TO SHIPMENT AND DAMAGE OF THIS MAGNITUDE WOULD HAVE BEEN DETECTED DURING THIS INSPECTION AND TESTING. OUR INSTRUCTION INSERT STATES: CARE SHOULD BE TAKEN NOT TO APPLY PRESSURE BETWEEN THE INSTRUMENT BLADE AND TISSUE PAD WITHOUT HAVING TISSUE BETWEEN THEM. THIS CAN RESULT IN POSSIBLE DAMAGE TO THE INSTRUMENT. BOTH CONDITIONS MAY CAUSE A SYSTEM FAILURE SIGNALED BY A CONTINUOUS BEEP WHEN EITHER OF THE FOOT PEDALS IS DEPRESSED AND KEEP THE CLAMP ARM OPEN WHEN BACKCUTTING OR WHILE THE BLADE IS ACTIVE WITHOUT TISSUE BETWEEN THE BLADE AND TISSUE PAD TO AVOID DAMAGE TO THE TISSUE PAD. WE HAVE DOCUMENTED THE CIRCUMSTANCES AS THEY WERE REPORTED TO US. IN ADDITION, COMPLAINT INFORMATION IS TRENDED ON A REGULAR BASIS TO DETERMINE IF FURTHER INVESTIGATION IS WARRANTED. THE BATCH HISTORY RECORDS WERE REVIEWED WITH NO ANOMALIES NOTED DURING THE MANUFACTURING PROCESS.

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING A LAP NISSEN FUNDOPLICATION, THE TISSUE PAD CAME OFF. THEY WERE ABLE TO RETRIEVE THE TISSUE PAD FROM THE PATIENT THROUGH THE TROCAR AND USED ANOTHER LIKE DEVICE TO COMPLETE THE CASE WITH NO PATIENT CONSEQUENCE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 ULTRACISION HARMONIC SCALPEL CURVED SHEARS (LCS) PISTOL GRIP 36CM - 15 MM ACTIVE GEI ETHICON ENDO-SURGERY, LLC. NA E4K26K

Patients

Seq Age Sex Outcome Treatment
1 GENERATOR| HANDPIECE