FDA Adverse Event Malfunction Summary report: N

ID NOW COVID-19

MDR report key: 10796312 · Received November 5, 2020

Report

Report Number
1221359-2020-00360
Event Type
Malfunction
Date Received
November 5, 2020
Date of Event
September 26, 2020
Report Date
December 8, 2020
Manufacturer
ABBOTT DIAGNOSTICS SCARBOROUGH, INC.
Product Code
QJR
UDI-DI
10811877011269
PMA / PMN Number
EUA2000074
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
IN, US
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

NEW INFORMATION WAS PROVIDED FROM ABBOTT DIAGNOSTICS (B)(4), INC.'S INVESTIGATION. A REVIEW OF THE COMPLAINTS REPORTED AS FALSE NEGATIVE PATIENT RESULTS (CONFIRMED AND UNCONFIRMED, CONFLICTING RESULTS) RELATED TO KIT LOT M126133 SHOWED THAT THE COMPLAINT RATE IS (B)(4).

Additional Manufacturer Narrative · 1

TESTING WAS PERFORMED AT ABBOTT DIAGNOSTICS (B)(4), INC. ON RETAINED KIT LOT M126133 WITH INTERNAL POSITIVE QUALITY CONTROL SAMPLES AND NEGATIVE QUALITY CONTROL SWABS. ALL TEST RESULTS WERE VALID AND PERFORMED AS EXPECTED. ADDITIONALLY, THE MANUFACTURING BATCH RECORDS AND QUALITY CONTROL RELEASE TESTING FOR KIT PART NUMBER 190-000 / LOT M126133 AND TEST BASE PART NUMBER 190-430 / LOT M126133 WERE REVIEWED. THIS LOT MET THE REQUIRED RELEASE SPECIFICATIONS. ABBOTT DIAGNOSTICS (B)(4), INC. WAS UNABLE TO DETERMINE THE EXACT ROOT CAUSE OF THE REPORTED ISSUE. THE AVAILABLE EVIDENCE SUGGESTS THAT THIS DEVICE LOT IS PERFORMING WITHIN THE STATEMENTS MADE WITHIN PACKAGE INSERT RELATED TO % TEST AGREEMENT WITH THE EXPECTED RESULTS BY SAMPLE CONCENTRATION.

Description of Event or Problem · 1

IT WAS REPORTED FROM CICU THAT (3) ALARIS PUMPS GAVE THE ERROR MESSAGE "CHANNEL COMMUNICATION ERROR" DURING LINE CHANGE, AND THERE WAS NO PATIENT IMPACT. PUMP ASSET TAG ID'S WERE CS1178, CS0393, CS0745, HOWEVER THE DEVICE SERIAL NUMBERS WERE NOT PROVIDED, AND THE PCU ID WAS UNKNOWN. ALTHOUGH REQUESTED, FURTHER INFORMATION WAS NOT PROVIDED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1255302 ID NOW COVID-19 MOLECULAR IVD FOR ID NOW COVID-19 QJR ABBOTT DIAGNOSTICS SCARBOROUGH, INC. M126133 10811877011269

Patients

Seq Age Sex Outcome Treatment
1