FDA Adverse Event Injury Summary report: N

ALBOGRAFT POLYESTER VASCULAR GRAFT

MDR report key: 10795871 · Received November 5, 2020

Report

Report Number
1220948-2020-00107
Event Type
Injury
Date Received
November 5, 2020
Date of Event
October 1, 2020
Report Date
November 5, 2020
Manufacturer
LEMAITRE VASCULAR, INC.
Product Code
MAL
UDI-DI
00840663102822
PMA / PMN Number
K093231
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
FR
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

WE COULD NOT PERFORM A HANDS-ON EVALUATION ON THE GRAFT SINCE IT HAS BEEN DISCARDED BY THE USER FACILITY. WE HAVE SENT A LIST OF FOLLOW-]UP QUESTIONS TO THE HOSPITAL BUT HAVE NOT YET RECEIVED A RESPONSE DESPITE OUR REPEATED ATTEMPTS. THE INVESTIGATION IS ONGOING, BUT OUR INQUIRIES ARE NOT GETTING ANY MORE RESPONSE DUE TO THE INCREASING COVID-19 SITUATION IN EUROPE. WE HAVE RELEASED A TOTAL OF (B)(4) UNITS OF ALBOGRAFT VASCULAR GRAFT FROM THIS LOT NUMBER. THIS GRAFT CATALOG# AMC6006; LOT# 211047 SERIAL# (B)(4) WAS SOLD TO THE HOSPITAL ON (B)(6) 2018. WE HAVE NOT RECEIVED ANY OTHER COMPLAINTS RELATED TO A SIMILAR INCIDENT FOR THIS LOT NUMBER. OUR REVIEW OF THE LOT HISTORY RECORDS FOR THIS GRAFT DID NOT REVEAL ANY ISSUES THAT COULD HAVE CONTRIBUTED TO THIS EVENT. WE HAVE ALSO REVIEWED OUR COMPLAINT HISTORY FOR LAST 5 YEARS. WE DID NOT FIND ANY OTHER COMPLAINTS RELATED TO A SIMILAR ISSUE REPORTED TO US BY OTHER HOSPITALS. PLEASE ALSO REFER TO MANUFACTURER INCIDENT REPORT# 1220948-2020-00106 RELATING TO ANOTHER CASE OF THROMBOSIS THAT OCCURRED AT THE SAME HOSPITAL AFTER IMPLANTING ALBOGRAFT VASCULAR GRAFT. WE HAVE RECEIVED A TOTAL OF (B)(4) COMPLAINTS FROM THIS HOSPITAL ON THE SAME DAY FOR THIS ISSUE WITH THE SAME EXACT DESCRIPTION WHICH RAISES QUESTIONS ON HOW THESE GRAFTS WERE PREPARED AND IMPLANTED, INCLUDING ANY POSTOPERATIVE CARE PROVIDED TO THE PATIENTS. WE HAVE ALSO REVIEWED OUR RISK DOCUMENT. THE CURRENT RATE OF OCCURRENCE FOR THIS ISSUE IS WITHIN OUR EXPECTED RATE. OUR IFU MENTIONS A LIST OF POTENTIAL COMPLICATIONS INCLUDING THROMBOSIS THAT OCCUR WITH THE USE OF ALBOGRAFT VASCULAR GRAFT.

Description of Event or Problem · 1

PROSTHESIS THROMBOSIS. FEM-POP ARTERIAL BYPASS. NEED OF ANOTHER SURGERY TO REPLACE IT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1256848 ALBOGRAFT POLYESTER VASCULAR GRAFT VASCULAR PROSTHESIS MAL LEMAITRE VASCULAR, INC. AMC3006 211047 00840663102822

Patients

Seq Age Sex Outcome Treatment
1 Required Intervention