SYNCHRON LXI 725 CLINICAL SYSTEM
Report
- Report Number
- 2050012-2008-00031
- Event Type
- Malfunction
- Date Received
- July 18, 2008
- Date of Event
- July 5, 2008
- Report Date
- July 18, 2008
- Manufacturer
- BECKMAN COULTER, INC.
- Product Code
- JJE
- PMA / PMN Number
- K023049
- Removal / Correction Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- MI, US
- Reporter Occupation
- NO INFORMATION
Narratives
THE SPECIMENS WERE COLLECTED IN LITHIUM HEPARIN TUBES. ON (B) (6), THE INSTRUMENT WAS GIVING OBSTRUCTION DETECTION ERRORS FOR THE CARTRIDGE CHEMISTRIES (CC) SAMPLE PROBE. THE CUSTOMER CALLED CUSTOMER TECHNICAL SERVICE (CTS), AND THEY ATTEMPTED TO RESOLVE THE ISSUE, BUT WERE UNABLE TO DO SO. THE OBSTRUCTION DETECTION FEATURE WAS DISABLED TO ALLOW THE CUSTOMER TO CONTINUE RUNNING CC. A FIELD SERVICE ENGINEER (FSE) WAS DISPATCHED TO THE CUSTOMER'S LAB ON (B) (6) 2008: THE FSE INSPECTED THE INSTRUMENT AND FOUND OBSTRUCTION IN THE CC SAMPLE PROBE. THE FSE REPLACED THE OBSTRUCTION DETECTION TRANSDUCER, THE CC SAMPLE PROBE AND, AS A PRECAUTION, THE 3-WAY VALVE. AFTER SERVICE, THE INSTRUMENT OPERATED WITHIN SPECIFICATIONS. ON A RECUR EVENT, ANY CARTRIDGE CHEMISTRY COULD BE AFFECTED. TREATMENT INITIATED OR WITHHELD BASED ON ERRONEOUS RESULTS COULD CONTRIBUTE TO SERIOUS INJURY. OBSTRUCTION IN THE CC SAMPLE PROBE MAY HAVE CONTRIBUTED TO THIS EVENT. A MALFUNCTION WILL BE ASSUMED FOR THE PURPOSE OF THIS REPORT. (B) (4).
A CUSTOMER CONTACTED BECKMAN COULTER INC. (BCI) REGARDING ERRONEOUS RESULTS FOR CARTRIDGE CHEMISTRIES (CC) GENERATED BY THE SYNCHRON LXI 725 CLINICAL SYSTEM. FOUR PT SAMPLES GAVE ERRONEOUS RESULTS FOR SEVERAL CHEMISTRIES. THE INCORRECT RESULTS WERE NOT REPORTED OUT OF THE LAB BECAUSE THEY DID NOT MATCH PREVIOUS PT VALUES. THE AFFECTED SAMPLES WERE RE-TESTED AND CORRECT RESULTS WERE OBTAINED. NO PT TREATMENTS WERE AFFECTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | SYNCHRON LXI 725 CLINICAL SYSTEM | CLINICAL CHEMISTRY ANALYZER | JJE | BECKMAN COULTER, INC. | LXI 725 | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | NA |