BD RAPID DETECTION OF SARS-COV-2 VERITOR ASSAY
Report
- Report Number
- 1119779-2020-00900
- Event Type
- Malfunction
- Date Received
- November 4, 2020
- Date of Event
- October 12, 2020
- Report Date
- February 25, 2021
- Manufacturer
- BECTON, DICKINSON & CO. (SPARKS)
- Product Code
- QKP
- PMA / PMN Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- TN, US
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
Narratives
EUA#: (B)(4). H6: INVESTIGATION SUMMARY BD HAS RECEIVED SEVERAL CUSTOMER COMPLAINTS FOR FALSE POSITIVE RESULTS, WHEN USING BD SARS-COV-2 REAGENTS FOR BD VERITOR¿ SYSTEM. THE CURRENT INVESTIGATION CONCERNS MULTIPLE LOTS OF BD SARS-COV-2 REAGENTS FOR BD VERITOR¿ SYSTEM. BD TAKES A SYSTEMATIC APPROACH TO INVESTIGATING FALSE POSITIVE COMPLAINTS THAT ARE RECEIVED. THIS APPROACH INVOLVES REVIEW OF MANUFACTURING BATCH HISTORY RECORDS, TESTING OF RETENTION SAMPLES, AND TESTING OF CUSTOMER RETURNED SAMPLES IF APPLICABLE. DUE TO UNKNOWN LOT# A DHR COULD NOT BE PERFORMED. THE INVESTIGATION DID NOT FIND A ROOT CAUSE FOR THE FALSE POSITIVE RESULTS THAT WERE OBSERVED. IT IS RECOMMENDED THAT EACH CUSTOMER REVIEW THEIR WORKFLOW CAREFULLY TO ENSURE THAT THE PACKAGE INSERT IS BEING FOLLOWED AS WRITTEN. THE ROOT CAUSE IS UNDER INVESTIGATION AND WILL BE DOCUMENTED IN OUR QUALITY SYSTEM. BD CANNOT CONFIRM THE COMPLAINT BASED ON THE INVESTIGATION THAT WAS PERFORMED. A CORRECTIVE AND PREVENTIVE ACTION (CAPA#1878253) IS ALREADY INITIATED TO INVESTIGATE THE ROOT CAUSE AND SOME MITIGATION ACTIONS ARE ALREADY BEING ADDRESSED. H3 OTHER TEXT : SEE H10.
IT WAS REPORTED WHILE TESTING FOR SARS COV-2 7 FALSE POSITIVE RESULTS WERE OBTAINED. REPEAT TESTS WERE PERFORMED USING PCR AND THE RESULTS WERE NEGATIVE. THE CUSTOMER STATED THEY ARE TESTING ASYMPTOMATIC PATIENTS. THIS TEST IS NOT INTENDED FOR USE ON ASYMPTOMATIC PATIENTS AND WAS THEREFORE USED OFF LABEL. THERE WAS NO INDICATION THAT RESULTS WERE REPORTED OUT AND THERE WAS NO REPORT OF PATIENT IMPACT. EUA#: (B)(4).
MEDICAL DEVICE EXPIRATION DATE: UNKNOWN. DEVICE MANUFACTURE DATE: UNKNOWN. A DEVICE EVALUATION AND/OR DEVICE HISTORY REVIEW IS ANTICIPATED, BUT IS NOT COMPLETE. UPON COMPLETION, A SUPPLEMENTAL REPORT WILL BE FILED. (B)(4).
IT WAS REPORTED WHILE TESTING FOR SARS COV-2 7 FALSE POSITIVE RESULTS WERE OBTAINED. REPEAT TESTS WERE PERFORMED USING PCR AND THE RESULTS WERE NEGATIVE. THE CUSTOMER STATED THEY ARE TESTING ASYMPTOMATIC PATIENTS. THIS TEST IS NOT INTENDED FOR USE ON ASYMPTOMATIC PATIENTS AND WAS THEREFORE USED OFF LABEL. THERE WAS NO INDICATION THAT RESULTS WERE REPORTED OUT AND THERE WAS NO REPORT OF PATIENT IMPACT. EUA#: (B)(4).
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1249656 | BD RAPID DETECTION OF SARS-COV-2 VERITOR ASSAY | CORONAVIRUS ANTIGEN DETECTION SYSTEM | QKP | BECTON, DICKINSON & CO. (SPARKS) | UNKNOWN |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |