SYRINGE 0.3ML 29GA 1/2IN 7BAG 420CAS JP
Report
- Report Number
- 1920898-2020-01524
- Event Type
- Malfunction
- Date Received
- November 4, 2020
- Date of Event
- October 14, 2020
- Report Date
- April 9, 2021
- Manufacturer
- BD MEDICAL - DIABETES CARE
- Product Code
- FMF
- PMA / PMN Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- JA
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
Narratives
H6: INVESTIGATION SUMMARY: CUSTOMER RETURNED PHOTOS OF 3/10CC, 12.7MM, 29G SYRINGES IN POLY BAGS FROM LOT # 0052894. CUSTOMER STATES THAT WHEN REMOVING THE SHIELD, THE NEEDLE WITH THE SHIELD WAS SEPARATED FROM THE BARREL. THE PHOTOS WERE EXAMINED AND EXHIBITED THE HUB-NEEDLE/SHIELD ASSEMBLY SEPARATED FROM THE BARREL. A REVIEW OF THE DEVICE HISTORY RECORD WAS COMPLETED FOR BATCH # 0052896 ALL INSPECTIONS WERE PERFORMED PER THE APPLICABLE OPERATIONS QC SPECIFICATIONS. THERE WAS ONE (1) NOTIFICATION NOTED FOR CRACKED HUBS CRACKED AND RAISED HUBS; BOTH DEFECTS MAY CAUSE THE NEEDLE ASSEMBLY UNIT TO NOT SNAP FIT ONTO THE BARREL. CAPA1630423 HAS BEEN OPENED TO ADDRESS THIS ISSUE. H3 OTHER TEXT : SEE H10.
IT WAS REPORTED THAT THE SYRINGE 0.3ML 29GA 1/2IN 7BAG 420CAS JP EXPERIENCED HUB SEPARATION FROM THE DEVICE. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: WHEN REMOVING THE SHIELD, THE NEEDLE WITH THE SHIED WAS SEPARATED FROM THE BARREL.
A DEVICE EVALUATION IS ANTICIPATED, BUT HAS NOT YET BEGUN. UPON COMPLETION OF THE INVESTIGATION, A SUPPLEMENTAL REPORT WILL BE FILED. FDA DEVICE PROBLEM CODE(S): 1354. FDA PATIENT PROBLEM CODE(S): 2199.
IT WAS REPORTED THAT THE SYRINGE 0.3ML 29GA 1/2IN 7BAG 420CAS JP EXPERIENCED HUB SEPARATION FROM THE DEVICE. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: WHEN REMOVING THE SHIELD, THE NEEDLE WITH THE SHIED WAS SEPARATED FROM THE BARREL.