CD HORIZON SPINAL SYSTEM
Report
- Report Number
- 1030489-2020-01583
- Event Type
- Malfunction
- Date Received
- November 4, 2020
- Date of Event
- October 8, 2020
- Report Date
- April 15, 2021
- Manufacturer
- MDT SOFAMOR DANEK PUERTO RICO MFG
- Product Code
- KWP
- PMA / PMN Number
- K202328
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- JA
- Reporter Occupation
- OTHER
Narratives
H3: PRODUCT ANALYSIS #704006274:PART # 5530030 LOT # H5578148 VISUAL MACROSCOPIC VISUAL AND MACROSCOPIC INSPECTION CONFIRMED THE THREADS OF THE SET SCREW HAVE BEEN DAMAGED. THE THREAD CREST AND FLANK DAMAGE APPEAR TO HAVE INITIATED AT THE START OF THE THREAD, AND IS CONSISTENT AROUND THE DAMAGED PORTION OF THE THREAD. THE FEMALE TORX OF THE SCREW HAS NOT BEEN DAMAGED. THIS TYPE OF DAMAGE IS CONSISTENT WITH MISALIGNMENT OF THE SET SCREW THREADS DURING CONSTRUCT ASSEMBLY. MEDTRONIC IS SUBMITTING THIS REPORT TO COMPLY WITH FDA REPORTING REGULATIONS UNDER 21 CFR PARTS 4 AND 803. THIS REPORT IS BASED UPON INFORMATION OBTAINED BY MEDTRONIC, WHICH THE COMPANY MAY NOT HAVE BEEN ABLE TO FULLY INVESTIGATE OR VERIFY PRIOR TO THE DATE THE REPORT WAS REQUIRED BY THE FDA. MEDTRONIC HAS MADE REASONABLE EFFORTS TO OBTAIN MORE COMPLETE INFORMATION AND HAS PROVIDED AS MUCH RELEVANT INFORMATION AS IS AVAILABLE TO THE COMPANY AS OF THE SUBMISSION DATE OF THIS REPORT. THIS REPORT DOES NOT CONSTITUTE AN ADMISSION OR A CONCLUSION BY FDA, MEDTRONIC, OR ITS EMPLOYEES THAT THE DEVICE, MEDTRONIC, OR ITS EMPLOYEE CAUSED OR CONTRIBUTED TO THE EVENT DESCRIBED IN THE REPORT. IN PARTICULAR, THIS REPORT DOES NOT CONSTITUTE AN ADMISSION BY ANYONE THAT THE PRODUCT DESCRIBED IN THIS REPORT HAS ANY ¿DEFECTS¿ OR HAS ¿MALFUNCTIONED¿. THESE WORDS ARE INCLUDED IN THE FDA 3500A FORM AND ARE FIXED ITEMS FOR SELECTION CREATED BY THE FDA TO CATEGORIZE THE TYPE OF EVENT SOLELY FOR THE PURPOSE OF REGULATORY REPORTING. MEDTRONIC OBJECTS TO THE USE OF THESE WORDS AND OTHERS LIKE THEM BECAUSE OF THE LACK OF DEFINITION AND THE CONNOTATIONS IMPLIED BY THESE TERMS. THIS STATEMENT SHOULD BE INCLUDED WITH ANY INFORMATION OR REPORT DISCLOSED TO THE PUBLIC UNDER THE FREEDOM OF INFORMATION ACT. ANY REQUIRED FIELDS THAT ARE UNPOPULATED ARE BLANK BECAUSE THE INFORMATION IS CURRENTLY UNKNOWN OR UNAVAILABLE. A GOOD FAITH EFFORT WILL BE MADE TO OBTAIN THE APPLICABLE INFORMATION RELEVANT TO THE REPORT. IF INFORMATION IS PROVIDED IN THE FUTURE, A SUPPLEMENTAL REPORT WILL BE ISSUED.
EVALUATION OF THE RETURNED DEVICE WAS NOT COMPLETED AT THE TIME OF THIS REPORT. A FOLLOW UP REPORT WILL BE SENT WHEN ANALYSIS IS COMPLETE. IF INFORMATION IS PROVIDED IN THE FUTURE, A SUPPLEMENTAL REPORT WILL BE ISSUED.
INFORMATION WAS RECEIVED FROM A HEALTHCARE PROVIDER (HCP) VIA MANUFACTURER REPRESENTATIVE REGARDING A PATIENT SPINAL THERAPY FOR PYO GENIC SPONDYLITIS. IT WAS REPORTED THAT THE REVISION SURGERY WAS PERFORMED FOR FUSION EXTENSION AND SET SCREW COULD NOT BE INSERTED. THERE WAS PATIENT INVOLVED IN THE EVENT AND NO FURTHER COMPLICATIONS WERE REPORTED. ADDITIONAL INFORMATION WAS RECEIVED ON 2020-OCT-15: THE PRODUCT WAS EXPLANTED ON (B)(6) 2020. THERE WAS NO PATIENT IMPACT ON PATIENT. ADDITIONAL INFORMATION WAS RECEIVED ON 2020-OCT-15: THE PRODUCTS E5 AND BA USED IN INITIAL SURGERY ARE MEDTRONIC AND NO MALFUNCTION ASSOCIATED WITH THEM. ADDITIONAL INFORMATION WAS RECEIVED ON 2020-OCT-21: MSB_UNK_SCREW WAS ADDED. REVISION SURGERY AND PYOGENIC SPONDYLITIS WAS REPORTED AFTER THE INITIAL SURGERY WHICH PERFORMED ON UNKNOWN DATE. THE SCREW SYSTEM IS LONGITUDE 5.5/6.0. NO FURTHER INFORMATION IS AVAILABLE. ADDITIONAL INFORMATION WAS RECEIVED ON 2020-NOV-04: THE REASON FOR REVISION SURGERY WAS FUSION FOR EXTENSION.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1251215 | CD HORIZON SPINAL SYSTEM | APPLIANCE, FIXATION, SPINAL INTERLAMINAL | KWP | MDT SOFAMOR DANEK PUERTO RICO MFG | 5530030 | H5578148 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |