Narratives

Additional Manufacturer Narrative · 1

THIS REPORTED EVENT AND SUBSEQUENT REPAIRS WERE INVESTIGATED THROUGH THE TSC TROUBLESHOOTING PROCESS. A REVIEW OF THE DEVICE SERVICE HISTORY RECORD WAS PERFORMED FROM THE DATE OF MANUFACTURE TO THE DATE CORRESPONDING TO THIS SERVICE NOTIFICATION NUMBER. THE DEVICE WAS RETURNED FOR SERVICE. THE DATABASE SHOWED NO QUALITY NOTIFICATIONS WERE OPENED FOR THE DEVICE. A REVIEW OF THE DEVICE HISTORY RECORD IN SAP FOR SN (B)(4) WAS PERFORMED FROM THE DATE OF THE MANUFACTURE TO DATE OF THE RELEASE OF PRODUCT, WHICH CONFIRMED THAT THIS DEVICE WAS NOT INVOLVED IN A PRODUCTION FAILURE, AND PRODUCT WAS RETURNED FOR SERVICING WHICH NOT CORRELATES TO THE CUSTOMER REPORTED ISSUE. A REVIEW OF THE COMPLAINT HISTORY RECORD IN THE TRACKWISE WAS PERFORMED FOR THE SN (B)(4) WHICH CONFIRMED NO SIMILAR COMPLAINTS WITH THE SAME OR RELATED FAILURE MODE. CAPA REFERENCE: (B)(4). THE CUSTOMER STATED THAT THERE WAS NO PATIENT INVOLVEMENT.

Description of Event or Problem · 1

(B)(4). 8100 13-1033-149 ERROR. RECOMMEND TO TRY TO REFLASH THE SOFTWARE. IF THAT DOES NOT WORK, LOGIC BOARD WILL HAVE TO BE REPLACED OR UNIT SENT IN FOR REPAIR. GAVE PRICE TO LEVEL 1 REPAIR. BIOMED WILL TRY TO REFLASH UNIT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1252285 8100 ALARIS PUMP MODULE PUMP, INFUSION FRN CAREFUSION SD 8100 10885403810015

Patients

Seq Age Sex Outcome Treatment
1