8015 ALARIS SYSTEM PC UNIT
Report
- Report Number
- 2016493-2020-28193
- Event Type
- Malfunction
- Date Received
- November 4, 2020
- Report Date
- April 3, 2018
- Manufacturer
- CAREFUSION SD
- Product Code
- FRN
- UDI-DI
- 10885403801518
- PMA / PMN Number
- K091308
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CA, US
- Reporter Occupation
- 003
Narratives
THIS REPORTED EVENT AND SUBSEQUENT REPAIRS WERE INVESTIGATED THROUGH THE SERVICE REPAIR PROCESS. FAILURE DATA AND PARTS-USED INFORMATION WERE REVIEWED FOR THE SAP AND TRACK WISE FILES AND FOUND RELEVANT TO THE SERVICE REPAIR. A REVIEW OF THE SOURCE DEVICE SERVICE HISTORY RECORD WAS PERFORMED FROM THE DATE OF MANUFACTURE TO THE DATE CORRESPONDING TO THIS SERVICE NOTIFICATION NUMBER. THE DATABASE SHOWED QUALITY NOTIFICATION # 200102941 WAS OPENED FOR THE DEVICE WITHOUT CORRELATION TO THE COMPLAINT. A REVIEW OF THE DEVICE HISTORY RECORD IN SAP FOR SN (B)(4) WAS PERFORMED FROM THE DATE OF THE MANUFACTURE TO DATE OF THE RELEASE OF PRODUCT, WHICH CONFIRMED THAT THIS DEVICE WAS INVOLVED IN A PRODUCTION FAILURE, AND PRODUCT WAS NOT RETURNED FOR SERVICING WHICH NOT CORRELATES TO THE CUSTOMER REPORTED ISSUE. A REVIEW OF THE COMPLAINT HISTORY RECORD IN THE TRACKWISE WAS PERFORMED FOR THE SN (B)(4) WHICH CONFIRMED NO SIMILAR COMPLAINTS WITH THE SAME OR RELATED FAILURE MODE. CAPA REFERENCE: (B)(4). THE CUSTOMER STATED THAT THERE WAS NO PATIENT INVOLVEMENT.
(B)(4). 8015 133.6090 ERROR. BIOMED REPLACED MAIN SPEAKER AND ERROR STILL CONTINUED. RECOMMEND TO TRY A SIO BOARD FROM A WORKING UNIT. IF ERROR GOES AWAY REPLACE SIO BOARD. IF ERROR STILL PRESENT REPLACE POWER SUPPLY PROCESSOR BOARD. GAVE PART NUMBER TO SIO BOARD AND POWER SUPPLY BOARD. BIOMED WILL TRY ANOTHER SIO BOARD.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1249377 | 8015 ALARIS SYSTEM PC UNIT | PUMP, INFUSION | FRN | CAREFUSION SD | 8015 | 10885403801518 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |