FDA Adverse Event Malfunction Summary report: N

SYRINGE 0.3ML 29GA 1/2IN 7BAG 420CAS JP

MDR report key: 10782497 · Received November 3, 2020

Report

Report Number
1920898-2020-01523
Event Type
Malfunction
Date Received
November 3, 2020
Date of Event
October 14, 2020
Report Date
March 3, 2021
Manufacturer
BD MEDICAL - DIABETES CARE
Product Code
FMF
PMA / PMN Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
JA
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 0

H.6. INVESTIGATION: NO SAMPLES WERE RETURNED THEREFORE THE INVESTIGATION WAS PERFORMED BASED ON THE PHOTOS PROVIDED. TWO PHOTOS OF A 0.3ML SYRINGE WERE PROVIDED. CONSUMER REPORTED WHEN REMOVING THE SHIELD, THE NEEDLE WITH THE SHIED WAS SEPARATED FROM THE BARREL. THE PHOTOS WERE REVIEWED, AND IT WAS OBSERVED THAT THE NEEDLE HUB/SHIELD ASSEMBLY WAS SEPARATED FROM THE BARREL; NO DAMAGE TO THE BARREL TIP WAS OBSERVED. DUE TO THE BATCH BEING UNKNOWN, NO DHR CAN BE PREFORMED. ROOT CAUSE FOR THIS DEFECT CANNOT BE DETERMINED. CAPA#1630423 WAS INITIATED.

Description of Event or Problem · 0

IT WAS REPORTED THAT THE SYRINGE 0.3ML 29GA 1/2IN 7BAG 420CAS JP EXPERIENCED HUB SEPARATION FROM THE DEVICE. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: WHEN REMOVING THE SHIELD, THE NEEDLE WITH THE SHIED WAS SEPARATED FROM THE BARREL.

Additional Manufacturer Narrative · 1

DEVICE EXPIRATION DATE: UNKNOWN. A DEVICE EVALUATION IS ANTICIPATED, BUT HAS NOT YET BEGUN. UPON COMPLETION OF THE INVESTIGATION, A SUPPLEMENTAL REPORT WILL BE FILED. DEVICE MANUFACTURE DATE: UNKNOWN. (B)(4).

Description of Event or Problem · 1

IT WAS REPORTED THAT THE SYRINGE 0.3ML 29GA 1/2IN 7BAG 420CAS (B)(4) EXPERIENCED HUB SEPARATION FROM THE DEVICE. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: WHEN REMOVING THE SHIELD, THE NEEDLE WITH THE SHIED WAS SEPARATED FROM THE BARREL.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1241146 SYRINGE 0.3ML 29GA 1/2IN 7BAG 420CAS JP SYRINGE FMF BD MEDICAL - DIABETES CARE UNKNOWN

Patients

Seq Age Sex Outcome Treatment
1 Other