SYRINGE 0.3ML 29GA 1/2IN 7BAG 420CAS JP
Report
- Report Number
- 1920898-2020-01510
- Event Type
- Malfunction
- Date Received
- November 2, 2020
- Date of Event
- October 13, 2020
- Report Date
- February 19, 2021
- Manufacturer
- BD MEDICAL - DIABETES CARE
- Product Code
- FMF
- PMA / PMN Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- JA
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
Narratives
H6: INVESTIGATION SUMMARY: CUSTOMER RETURNED PHOTOS OF A 3/10CC SYRINGE. CUSTOMER STATES THAT THE HUB-NEEDLE/SHIELD ASSEMBLY WAS SEPARATED FROM THE BARREL. THE PHOTOS WERE EXAMINED AND EXHIBITED THE HUB-NEEDLE/SHIELD ASSEMBLY SEPARATED FROM THE BARREL. UNABLE TO PERFORM DHR CHECK FOR NEEDLE HUB SEPARATES DUE TO UNKNOWN LOT NUMBER. BD WAS ABLE CONFIRM THE CUSTOMER¿S INDICATED FAILURE. ROOT CAUSE FOR THIS DEFECT CANNOT BE DETERMINED. CAPA1630423 HAS BEEN OPENED TO ADDRESS THIS ISSUE.
IT WAS REPORTED THAT THE SYRINGE 0.3ML 29GA 1/2IN 7BAG 420CAS JP EXPERIENCED HUB SEPARATION FROM THE DEVICE. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: WHEN REMOVING THE SHIELD BEFORE USE, THE NEEDLE WITH THE SHIED WAS SEPARATED FROM THE BARREL.
A DEVICE EVALUATION IS ANTICIPATED, BUT HAS NOT YET BEGUN. UPON COMPLETION OF THE INVESTIGATION, A SUPPLEMENTAL REPORT WILL BE FILED. (B)(4).
IT WAS REPORTED THAT THE SYRINGE 0.3ML 29GA 1/2IN 7BAG 420CAS JP EXPERIENCED HUB SEPARATION FROM THE DEVICE. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: WHEN REMOVING THE SHIELD BEFORE USE, THE NEEDLE WITH THE SHIED WAS SEPARATED FROM THE BARREL.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1241043 | SYRINGE 0.3ML 29GA 1/2IN 7BAG 420CAS JP | SYRINGE | FMF | BD MEDICAL - DIABETES CARE | UNKNOWN |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |