BD SEDI-40
Report
- Report Number
- 2243072-2020-01771
- Event Type
- Malfunction
- Date Received
- November 2, 2020
- Date of Event
- October 8, 2020
- Report Date
- December 11, 2020
- Manufacturer
- BECTON DICKINSON
- Product Code
- JPH
- PMA / PMN Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- RO
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
Narratives
THE FOLLOWING FIELDS WERE UPDATED DUE TO ADDITIONAL INFORMATION: D10: DEVICE AVAILABLE FOR EVAL YES. D10: RETURNED TO MANUFACTURER ON: 2020-11-09. H6: INVESTIGATION SUMMARY: INSTRUMENT SEDI 40 00203 WAS RETURNED TO THE MANUFACTURER FOR SERVICE WITH RESPECT TO THE REPORTED DEFECT ¿ NOT MIXING. THE INSTRUMENT WAS EVALUATED BY VISUAL EXAMINATION AND FUNCTIONAL TESTING AND TWO LOOSE SCREWS WERE FOUND, WITHIN THE INSTRUMENT, THAT HOLDS THE TUBE HOLDER. THE SCREWS WERE RESEATED. BASED ON A REVIEW OF THE DEVICE HISTORY RECORD FOR THE INCIDENT LOT, ALL PRODUCT SPECIFICATIONS AND REQUIREMENTS FOR LOT RELEASE WERE MET. THERE WERE NO RELATED QUALITY NON-CONFORMANCES DURING MANUFACTURING OF THE PRODUCT.
IT WAS REPORTED THAT DURING USE WITH A BD SEDI-40 THE MIXER IS NOT WORKING. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: MIXER DOES NOT WORK ANY MORE.
FOR OEM MANUFACTURING SITES: (B)(4). MEDICAL DEVICE EXPIRATION DATE: NA. A DEVICE EVALUATION IS ANTICIPATED, BUT HAS NOT YET BEGUN. UPON COMPLETION OF THE INVESTIGATION, A SUPPLEMENTAL REPORT WILL BE FILED. DEVICE MANUFACTURE DATE: UNKNOWN. (B)(4).
IT WAS REPORTED THAT DURING USE WITH A BD SEDI-40 THE MIXER IS NOT WORKING. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: MIXER DOES NOT WORK ANY MORE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1234617 | BD SEDI-40 | ESR INSTRUMENT | JPH | BECTON DICKINSON | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |