FDA Adverse Event Malfunction Summary report: N

CAPSTONE SPINAL SYSTEM

MDR report key: 10769522 · Received November 2, 2020

Report

Report Number
1030489-2020-01548
Event Type
Malfunction
Date Received
November 2, 2020
Date of Event
October 12, 2020
Report Date
March 31, 2021
Manufacturer
MDT SOFAMOR DANEK PUERTO RICO MFG
Product Code
MAX
UDI-DI
00613994290991
PMA / PMN Number
K073291
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
TW
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 0

H3: PRODUCT ANALYSIS: PART # 2991026, LOT # H5603469. VISUAL AND OPTICAL INSPECTION CONFIRMED THE THREADED ATTACHMENT PORTION OF THE CAGE HAS BROKEN AWAY. THE THREADS HAVE BEEN DAMAGED AND STRIPPED. THE INSERTER MAY HAVE BEEN SLIGHT BENT WHEN BACKED OUT FROM THE IMPLANT CAUSING THE CAGE TO BREAK. UNABLE TO DETERMINE ROOT CAUSE OF BREAK. MEDTRONIC IS SUBMITTING THIS REPORT TO COMPLY WITH FDA REPORTING REGULATIONS UNDER 21 CFR PARTS 4 AND 803. THIS REPORT IS BASED UPON INFORMATION OBTAINED BY MEDTRONIC, WHICH THE COMPANY MAY NOT HAVE BEEN ABLE TO FULLY INVESTIGATE OR VERIFY PRIOR TO THE DATE THE REPORT WAS REQUIRED BY THE FDA. MEDTRONIC HAS MADE REASONABLE EFFORTS TO OBTAIN MORE COMPLETE INFORMATION AND HAS PROVIDED AS MUCH RELEVANT INFORMATION AS IS AVAILABLE TO THE COMPANY AS OF THE SUBMISSION DATE OF THIS REPORT. THIS REPORT DOES NOT CONSTITUTE AN ADMISSION OR A CONCLUSION BY FDA, MEDTRONIC, OR ITS EMPLOYEES THAT THE DEVICE, MEDTRONIC, OR ITS EMPLOYEE CAUSED OR CONTRIBUTED TO THE EVENT DESCRIBED IN THE REPORT. IN PARTICULAR, THIS REPORT DOES NOT CONSTITUTE AN ADMISSION BY ANYONE THAT THE PRODUCT DESCRIBED IN THIS REPORT HAS ANY ¿DEFECTS¿ OR HAS ¿MALFUNCTIONED¿. THESE WORDS ARE INCLUDED IN THE FDA 3500A FORM AND ARE FIXED ITEMS FOR SELECTION CREATED BY THE FDA TO CATEGORIZE THE TYPE OF EVENT SOLELY FOR THE PURPOSE OF REGULATORY REPORTING. MEDTRONIC OBJECTS TO THE USE OF THESE WORDS AND OTHERS LIKE THEM BECAUSE OF THE LACK OF DEFINITION AND THE CONNOTATIONS IMPLIED BY THESE TERMS. THIS STATEMENT SHOULD BE INCLUDED WITH ANY INFORMATION OR REPORT DISCLOSED TO THE PUBLIC UNDER THE FREEDOM OF INFORMATION ACT. ANY REQUIRED FIELDS THAT ARE UNPOPULATED ARE BLANK BECAUSE THE INFORMATION IS CURRENTLY UNKNOWN OR UNAVAILABLE. A GOOD FAITH EFFORT WILL BE MADE TO OBTAIN THE APPLICABLE INFORMATION RELEVANT TO THE REPORT. IF INFORMATION IS PROVIDED IN THE FUTURE, A SUPPLEMENTAL REPORT WILL BE ISSUED.

Additional Manufacturer Narrative · 1

IF INFORMATION IS PROVIDED IN THE FUTURE, A SUPPLEMENTAL REPORT WILL BE ISSUED.

Description of Event or Problem · 1

INFORMATION WAS RECEIVED FROM HEALTHCARE PROFESSIONAL VIA MANUFACTURER REPRESENTATIVE REGARDING A PATIENT HAVING TLIF AT L5/S1 DUE TO L5/S1 SPONDYLOSIS AND FORAMINAL STENOSIS. IT WAS REPORTED THAT CAGE'S PART THAT INSERTER HOLD HAD BROKEN AND REMAINED ON THE INSERTER WHEN THE INSERTER WAS TAKEN OFF THE BODY. THE CAGE REMAINS IMPLANTED. DATE OF IMPLANT IS (B)(6) 2020. THERE WERE NO PATIENT SYMPTOMS/ COMPLICATIONS AS A RESULT OF THE EVENT. NO FURTHER COMPLICATIONS WERE REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1236160 CAPSTONE SPINAL SYSTEM INTERVERTEBRAL FUSION DEVICE WITH BONE GRAFT, LUMBAR MAX MDT SOFAMOR DANEK PUERTO RICO MFG 2991026 H5603469 00613994290991

Patients

Seq Age Sex Outcome Treatment
1 34 YR