FDA Adverse Event Malfunction Summary report: N

ENDOWRIST;DAVINCI SI

MDR report key: 10767663 · Received November 1, 2020

Report

Report Number
2955842-2020-11148
Event Type
Malfunction
Date Received
November 1, 2020
Date of Event
September 3, 2020
Report Date
October 7, 2020
Manufacturer
INTUITIVE SURGICAL, INC
Product Code
NAY
UDI-DI
00886874111642
PMA / PMN Number
K050369
Removal / Correction Number
N/A
Product Problem
Yes
Report Source
Manufacturer report
Reporter Occupation
003

Narratives

Additional Manufacturer Narrative · 1

INTUITIVE SURGICAL, INC. (ISI) RECEIVED THE FENESTRATED BIPOLAR FORCEPS INSTRUMENT INVOLVED WITH THIS COMPLAINT, AND COMPLETED THE DEVICE EVALUATION. THE CUSTOMER REPORTED EVENT OF NON-INTUITIVE MOTION WAS NOT REPRODUCED THROUGH FAILURE ANALYSIS INVESTIGATION. THE INSTRUMENT PASSED THE RECOGNITION AND ENGAGEMENT TESTS, AND MOVED INTUITIVELY WITH FULL RANGE OF MOTION IN ALL DIRECTIONS DURING TESTING. THE INSTRUMENT GRIPS OPENED AND CLOSED PROPERLY. AS PART OF INVESTIGATION, INSPECTION OF THE INSTRUMENT FOUND INSULATION DAMAGE ON THE CONDUCTOR WIRE. NO THERMAL DAMAGE WAS IDENTIFIED. THE INSTRUMENT WAS TESTED FOR ELECTRICAL CONTINUITY, AND PASSED. SITE HISTORY REVIEW: A REVIEW OF THE SITE'S COMPLAINT HISTORY DOES NOT SHOW ANY ADDITIONAL COMPLAINTS RELATED TO THIS PRODUCT. NO PROCEDURE VIDEO OR IMAGE WAS SUBMITTED TO ISI FOR REVIEW, NO ADDITIONAL TECHNICAL ANALYSIS WAS CONDUCTED. SYSTEM LOG INVESTIGATION: A REVIEW OF THE INSTRUMENT LOG FOR THE FENESTRATED BIPOLAR FORCEPS INSTRUMENT LOT# N10190624 / SEQUENCE 961 ASSOCIATED WITH THIS EVENT HAS BEEN PERFORMED. PER LOGS, THE INSTRUMENT WAS LAST USED ON (B)(6) 2020 USING SYSTEM SH2264 AND NOT ON THE REPORTED EVENT DATE OF (B)(6) 2020. THE INSTRUMENT HAS 3 REMAINING USABLE LIVES WITH NO SUBSEQUENT USE RECORDED AFTER (B)(6) 2020. ADDITIONALLY, TWO PROCEDURES WERE LOGGED ON THE REPORTED EVENT DATE OF (B)(6) 2020. HOWEVER, NO USAGE OF A FENESTRATED BIPOLAR FORCEPS INSTRUMENT WAS LOGGED ON EITHER OF THE PROCEDURES. THIS COMPLAINT IS BEING REPORTED AS A REPORTABLE MALFUNCTION EVENT DUE TO THE FOLLOWING CONCLUSION: THE FENESTRATED BIPOLAR FORCEPS INSTRUMENTS ARE MULTIPLE-USE ELECTROSURGICAL ENDOSCOPIC INSTRUMENTS WITH A GRASPING TIP TO BE USED IN CONJUNCTION WITH THE DA VINCI SYSTEM AND AN EXTERNAL ELECTROSURGICAL UNIT (ESU). THE INSTRUMENT IS DESIGNED TO PROVIDE ENERGY FROM THE DESIGNATED LOCATION ON THE INSTRUMENT (THE TIP) TO THE PLANNED ANATOMICAL LOCATION WHEN USED AS INTENDED. THE ENERGY IS ACTIVATED BY PRESSING THE DESIGNATED PEDAL ON THE SURGEON SIDE CONSOLE (SSC). IT WAS REPORTED THAT DURING A DA VINCI-ASSISTED SURGICAL PROCEDURE, THE FENESTRATED BIPOLAR FORCEPS INSTRUMENT EXHIBITED NON-INTUITIVE MOTION. THE USER COMPLETED THE PROCEDURE WITH NO FURTHER ISSUE REPORTED. THE CUSTOMER REPORTED EVENT OF NON-INTUITIVE MOTION WAS NOT REPRODUCED THROUGH FAILURE ANALYSIS INVESTIGATION. ADDITIONAL OBSERVATION DURING INVESTIGATION IDENTIFIED A COMPROMISED CONDUCTOR WIRE INSULATION WITH A PASSED ELECTRICAL CONTINUITY TEST. THE CUSTOMER REPORTED COMPLAINT DOES NOT ITSELF CONSTITUTE AN MDR REPORTABLE EVENT, HOWEVER, THIS COMPLAINT IS BEING REPORTED BECAUSE DAMAGE TO THE CONDUCTOR WIRE WITHIN THE INSTRUMENT WITH A PASSED ELECTRICAL CONTINUITY TEST FOUND DURING FAILURE ANALYSIS COULD LEAD TO UNINTENDED ELECTRICAL DISCHARGE AT A LOCATION OTHER THAN INTENDED. ALTHOUGH THERE WAS NO ARCING REPORTED, OR SEEN THROUGH ANALYSIS, AND THERE WAS NO PATIENT INJURY REPORTED, IF THIS FAILURE WERE TO RECUR, IT COULD RESULT IN AN ADVERSE EVENT. DEVICE EXPIRATION DATE WAS LEFT BLANK AS THIS INSTRUMENT HAS 10 USAGES ALLOTTED TO IT, WHICH ARE TRACKED BY THE DA VINCI SURGICAL SYSTEM. THE INSTRUMENT HAS REMAINING USABLE LIVES, AND THEREFORE, HAD NOT EXPIRED. THE PRODUCT IS NOT IMPLANTABLE.

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING A DA VINCI-ASSISTED SURGICAL PROCEDURE, THE FENESTRATED BIPOLAR FORCEPS INSTRUMENT EXHIBITED NON-INTUITIVE MOTION. THE USER COMPLETED THE PROCEDURE WITH NO FURTHER ISSUE REPORTED. NO KNOWN IMPACT, OR PATIENT CONSEQUENCE WAS REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1233103 ENDOWRIST;DAVINCI SI FENESTRATED BIPOLAR FORCEPS NAY INTUITIVE SURGICAL, INC 420205-15 N10190624 961 00886874111642

Patients

Seq Age Sex Outcome Treatment
1