ALARIS PUMP MODULE
Report
- Report Number
- 2016493-2020-30227
- Event Type
- Malfunction
- Date Received
- October 30, 2020
- Report Date
- October 7, 2020
- Manufacturer
- CAREFUSION SD
- Product Code
- FRN
- UDI-DI
- 10885403810015
- PMA / PMN Number
- K133532
- Removal / Correction Number
- Z-2822-2020
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- NC, US
- Reporter Occupation
- BIOMEDICAL ENGINEER
Narratives
THE REPORTED ISSUE OF A DIM SEGMENT IS CONFIRMED BASED ON THE FIELD ACTION. NO FURTHER COMPLAINT INVESTIGATION IS NECESSARY AS CAPA: 1143616 WAS OPENED TO ADDRESS THIS ISSUE. DEVICE HISTORY RECORD (DHR) REVIEWS ARE NOT REQUIRED FOR FIELD ACTION COMPLAINTS BECAUSE THE ROOT CAUSE OF THE ISSUE IS KNOWN, THE INVESTIGATION IS DOCUMENTED WITHIN A CAPA, AND A FIELD ACTION HAS BEEN INITIATED. THE DEVICE IS USED FOR TREATMENT PURPOSES.
IT WAS REPORTED THAT ALLEGEDLY THE DISPLAY SEGMENT WAS DIM FOR ONE LARGE VOLUME PUMP MODULES, AND THEREFORE, WILL BE REPLACED.
NO DEVICE WILL BE RETURNED PER CUSTOMER. THE CUSTOMER COMPLAINT COULD NOT BE CONFIRMED BECAUSE THE DEVICE WAS NOT RETURNED FOR FAILURE INVESTIGATION. THE ROOT CAUSE OF THIS FAILURE WAS NOT IDENTIFIED.
IT WAS REPORTED THAT ALLEGEDLY THE DISPLAY SEGMENT WAS DIM FOR ONE LARGE VOLUME PUMP MODULES, AND THEREFORE, WILL BE REPLACED.